Key result
Multicenter flow-mediated dilatation demonstrates reproducibility with ~1% mean absolute difference over 48 hours.
Observational (n=67)
Yes
Randomized trial assesses reproducibility of flow-mediated dilatation in multicentre setting, indicating its reliability for evaluations.
AIMS: The aim of this study was to assess the reproducibility of flow-mediated dilatation (FMD) in a multicentre setting. METHODS AND RESULTS: This study was performed as part of the dal-VESSEL trial in which FMD was measured in 19 vascular imaging centres in six European countries. A subgroup of patients who were allocated in the placebo group and scanned twice at each trial time point (substudy) was analysed. Intra-sonographer variability was calculated from FMD measurements 48 h apart. Centre variability and short-, medium-, and long-term reproducibility of FMD were calculated at 48 h and at 3 and 9 months intervals, respectively. Intra- and inter-reader variability was assessed by re-analysing the FMD images by three certified readers at two time intervals, 7 days apart. Sixty-seven patients were included. Variability between centres was comparable at 48 h and 3 months interval but almost doubled at 9 months. The mean absolute difference in %FMD was 1.04, 0.99, and 1.45% at the three time intervals, respectively. Curves were generated to indicate the number of patients required for adequate power in crossover and parallel study designs. CONCLUSION: This study demonstrates for the first time that in a multicentre setting reproducible FMD measurements can be achieved for short- and medium-term evaluation, which are comparable with those reported from specialized laboratories. These findings justify the use of FMD as an outcome measure for short- and medium-term assessment of pharmacological interventions.
No takes yet. Share an insight, caveat, or question.
Charakida et al. (2013) reported an observational. Flow-mediated dilatation (FMD) measurement was evaluated on Mean absolute difference in %FMD at 48 hours, 3 months, and 9 months. Flow-mediated dilatation measurements showed reproducibility with mean absolute differences in %FMD of 1.04% at 48 hours, 0.99% at 3 months, and 1.45% at 9 months in a multicenter setting.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: