Key result
The Apple Watch Series 4 successfully obtained 3-lead ECG tracings with similar duration, amplitude, and polarity compared to standard ECG across 300 tracings.
Why the study?
The reliability of obtaining 3-lead ECG recordings with smartwatches has yet to be confirmed in a large study.
Does a smartwatch reliably obtain 3-lead ECG recordings compared to standard ECG?
Observational
Does a smartwatch reliably obtain 3-lead ECG recordings compared to standard ECG?
The Apple Watch Series 4 can feasibly and reliably obtain 3-lead ECG tracings (Einthoven leads I, II, and III) that are comparable to standard ECG.
Supports smartwatch 3-lead ECG feasibility; leaves open prospective validation before clinical adoption.
Some of the recently released smartwatch products feature a single-lead electrocardiogram (ECG) recording capability. The reliability of obtaining 3-lead ECG with smartwatches is yet to be confirmed in a large study. This study aimed to assess the feasibility and reliability of smartwatch to obtain 3-lead ECG recordings, the classical Einthoven ECG leads I-III compared to standard ECG. To record lead I, the watch was worn on the left wrist and the right index finger was placed on the digital crown for 30 s. For lead II, the watch was placed on the lower abdomen and the right index finger was placed on the digital crown for 30 s. For lead III, the same process was repeated with the left index finger. Spearman correlation and Bland-Altman tests were used for data analysis. A total of 300 smartwatch ECG tracings were successfully obtained. ECG waves' characteristics of all three leads obtained from the smartwatch had a similar duration, amplitude, and polarity compared to standard ECG. The results of this study suggested that the examined smartwatch (Apple Watch Series 4) could obtain 3-lead ECG tracings, including Einthoven leads I, II, and III by placing the smartwatch on the described positions.
No takes yet. Share an insight, caveat, or question.
Behzadi et al. (2020) reported an observational. Smartwatch (Apple Watch Series 4) vs. Standard ECG was evaluated on ECG waves' characteristics (duration, amplitude, and polarity). The Apple Watch Series 4 successfully obtained 3-lead ECG tracings with similar duration, amplitude, and polarity compared to standard ECG across 300 tracings.