Key result
In patients undergoing CRT-D implantation, vector reprogramming using a bipolar LV lead successfully avoided phrenic nerve stimulation in 97.5% of patients, exceeding the 90% performance goal.
Why the study?
Does vector reprogramming with a bipolar LV lead prevent phrenic nerve stimulation in patients undergoing CRT-D implantation?
Observational (n=90)
Yes
Does vector reprogramming with a bipolar LV lead prevent phrenic nerve stimulation in patients undergoing CRT-D implantation?
Bipolar LV leads with vector reprogramming can successfully avoid phrenic nerve stimulation in 97.5% of patients during CRT-D implantation, offering a reliable alternative when coronary sinus anatomy precludes quadripolar leads.
Answers are grounded in this paper and cite their sources.
Bipolar LV leads with vector reprogramming may manage PNS during CRT-D implantation; leaves open their role versus quadripolar leads in routine practice.
Background: Up to one in three patients implanted with a cardiac resynchronization therapy-defibrillator (CRT-D) device experience phrenic nerve stimulation (PNS). Quadripolar leads are effective at reducing PNS, but compared to standard bipolar leads they have limitations related to maneuverability and high pacing thresholds. The ability of standard bipolar leads to overcome PNS is explored here. Methods: The French multicenter, observational study ORPHEE enrolled 90 CRT-D-eligible patients. Detection of PNS took place after satisfactory positioning of the LV bipolar lead (stable pacing threshold - LV ring, LV tip - RV ring and LV ring - RV coil) could prevent PNS from occurring in at least 90% of patients. Results: In 80 evaluable patients, PNS was reported in 12 patients (15%). Reprogramming overcame PNS in 10 patients: LV ring - RV coil in 8 patients; LV tip - LV ring in 1; and LV tip- RV ring in 1. As PNS was avoided in 78 of 80 patients (97.5%), the primary endpoint was significant (97.5% vs. 90%, p = 0.01). Conclusion: During CRT-D implantation, PNS occurred in 15% of patients. In most (97.5%) implanted patients, PNS could be avoided by vector reprogramming using a bipolar LV lead. For patients whose coronary sinus anatomy precludes the implantation of multi-electrode leads, bipolar leads are a suitable, reliable alternative.
No takes yet. Share an insight, caveat, or question.
Benkemoun et al. (2017) conducted an observational in Heart failure eligible for CRT-D (n=90). Bipolar left ventricular leads with three programmable pacing vectors vs. 90% pre-specified performance goal was evaluated on Prevention or resolution of phrenic nerve stimulation using three programmable pacing vectors at implant (p=0.01). In patients undergoing CRT-D implantation, vector reprogramming using a bipolar LV lead successfully avoided phrenic nerve stimulation in 97.5% of patients, exceeding the 90% performance goal.