Key result
Estimating carotid-femoral pulse wave velocity from finger photoplethysmography yielded acceptable discrepancies compared to Doppler ultrasound.
Why the study?
This project aimed to compare a new method for estimating carotid-femoral pulse wave velocity from finger photoplethysmography with the gold-standard technique.
Can carotid-femoral pulse wave velocity be accurately estimated from finger photoplethysmography signals compared to Doppler ultrasound?
Cross-Sectional (n=90)
Can carotid-femoral pulse wave velocity be accurately estimated from finger photoplethysmography signals compared to Doppler ultrasound?
Carotid-femoral pulse wave velocity can be excellently estimated by processing finger photoplethysmography signals along with demographic and hemodynamic variables using an Exponential Gaussian process regression model.
Answers are grounded in this paper and cite their sources.
Finger photoplethysmography shows acceptable cf-PWV agreement in this cohort; leaves open larger prospective validation before clinical adoption.
Objective . This project compared a new method to estimate the carotid-femoral pulse wave velocity (cf-PWV) to the gold-standard cf-PWV technique. Approach . The cf-PWV was estimated from the pulse transit time (FPS-PTT) calculated by processing the finger photoplethysmographic signal of Finapres (FPS) and subject’s height only (brief mode) as well as along with other variables (age, heart rate, arterial pressure, weight; complete mode). Doppler ultrasound cf-PWVs and FPS-PTTs were measured in 90 participants equally divided into 3 groups (18–30; 31–59; 60–79 years). Predictions were performed using multiple linear regressions (MLR) and with the best regression model identified by using MATLAB Regression Learner App. A validation set approach (60 training datasets, 30 testing datasets; VSA) and leave-one-out cross-validation (LOOCV) were used. Main results . With MLR, the discrepancies were: 0.01 ± 1.21 m s −1 (VSA) and 0.001 ± 1.11 m s −1 (LOOCV) in brief mode; −0.02 ± 0.83 m s −1 (VSA) and 0.001 ± 0.84 m s −1 (LOOCV) in complete mode. Using a linear support vector machine model (SVM) in brief mode, the discrepancies were: 0.01 ± 1.19 m s −1 (VSA) and −0.01 ± 1.06 m s −1 (LOOCV). Using an Exponential Gaussian process regression model (GPR) in complete mode, the discrepancies were: −0.03 ± 0.79 m s −1 (VSA) and 0.01 ± 0.75 m s −1 (LOOCV). Significance . The cf-PWV can be estimated by processing the FPS-PTT and subjects’ height only, but the inclusion of other variables improves the prediction performance. Predictions through MLR qualify as acceptable in both brief and complete modes. Predictions via linear SVM in brief mode improve but still qualify as acceptable. Interestingly, predictions through Exponential GPR in complete mode improve and qualify as excellent.
No takes yet. Share an insight, caveat, or question.
Gentilin et al. (2022) reported a cross-sectional. Finger photoplethysmography signal (FPS-PTT) estimation vs. Doppler ultrasound cf-PWV was evaluated on Discrepancy between estimated and gold-standard cf-PWV. Estimating carotid-femoral pulse wave velocity from finger photoplethysmography yielded acceptable discrepancies compared to Doppler ultrasound.