Key result
The Biophen assay showed agreement with LC-MS/MS in non-bleeding patients but overestimated apixaban and rivaroxaban concentrations by 26% in spiked plasma from patients with active bleeding.
Why the study?
Does the Biophen DiXaI chromogenic assay accurately measure apixaban and rivaroxaban concentrations compared to LC-MS/MS in bleeding and non-bleeding patients?
Observational (n=120)
Does the Biophen DiXaI chromogenic assay accurately measure apixaban and rivaroxaban concentrations compared to LC-MS/MS in bleeding and non-bleeding patients?
The Biophen chromogenic assay is suitable for measuring apixaban and rivaroxaban in stable patients but overestimates concentrations by 26% in actively bleeding patients, which must be considered during intoxication assessment.
Answers are grounded in this paper and cite their sources.
Biophen DiXaI overestimates DOAC levels 26% in bleeding; leaves open prospective validation before guiding management.
Objectives The direct oral anticoagulants ( DOAC s), apixaban and rivaroxaban, are used for anticoagulation treatment. Although biochemical monitoring is not required, severe bleedings caused by DOAC have been reported. We therefore evaluated the chromogenic assay Biophen DiXaI ® (Biophen) using plasma from non‐bleeding patients in treatment with DOAC and on spiked plasma from patients with ongoing bleeding. Methods The Biophen method was compared with an in‐house reference method; liquid chromatography‐tandem mass spectrometry ( LC ‐ MS / MS ) using plasma spiked with the compounds ranging from 20 to 1500 μg/L. Furthermore, the methods were compared using plasma from non‐bleeding patients in treatment with DOAC (n = 106). In addition, plasma was collected from patients not treated with DOAC , but with ongoing bleeding and tested in spiking experiments (n = 14). Results Analysis of plasma from 106 patients receiving rivaroxaban (n = 91) or apixaban (n = 15) showed agreement and correlation between the methods. Measurement in spiked plasma from patients with active bleeding, however, revealed a 26% overestimation by the assay. Conclusions Our findings show that Biophen is suitable for measuring apixaban and rivaroxaban concentrations in plasma. However, our results in patients with active bleeding show an overestimation of 26%. This should be taken into account when assessing possible intoxication with apixaban and rivaroxaban in patients presenting with bleeding.
No takes yet. Share an insight, caveat, or question.
Jensen et al. (2017) conducted an observational in DOAC treatment and active bleeding (n=120). Biophen DiXaI chromogenic assay vs. Liquid chromatography-tandem mass spectrometry (LC-MS/MS) was evaluated on Measurement of apixaban and rivaroxaban concentrations (26% overestimation). The Biophen assay showed agreement with LC-MS/MS in non-bleeding patients but overestimated apixaban and rivaroxaban concentrations by 26% in spiked plasma from patients with active bleeding.
Synapse has enriched 10 related papers. Consider them for comparative context: