Key result
A multiplanar transesophageal ultrasonic imaging system accurately measured the volume of water-filled balloons and excised hearts in vitro.
Why the study?
Does a multiplanar transesophageal ultrasonic imaging system accurately measure ventricular volume in vitro?
Does a multiplanar transesophageal ultrasonic imaging system accurately measure ventricular volume in vitro?
A novel multiplanar transesophageal ultrasonic imaging system demonstrates high accuracy for measuring ventricular volume in vitro.
Answers are grounded in this paper and cite their sources.
Supports in vivo validation of multiplanar TEE volumetry; leaves open clinical translation.
We have developed a system to assess the feasibility of using multiple transesophageal ultrasonic images to measure left-ventricular volume, an important variable in patient management. The system includes a special transesophageal probe with a micromanipulator for acquiring cardiac images in multiple planes with known interplanar spatial relationship and an off-line processing system to compute the volume. In vitro studies with the probe demonstrated that the distance between two targets in space can be identified within 2 mm (SD = 0.4 mm) for points in the imaging plane 3.4 mm (SD = 0.5 mm) for points not lying in the imaging plane. This gives an average accuracy of +/- 6.5% for distances greater than 4.5 cm. Comparison of ultrasonic measurements of the volume of water-filled balloons and excised hearts to the volume required to fill them, revealed a correlation coefficient of 0.992, a regression line having a slope of 1.0 and an ordinate intercept at 0.2 mL, and a standard error of the estimate of 8 mL.
No takes yet. Share an insight, caveat, or question.
Martin et al. (1990) studied this question. Multiplanar transesophageal ultrasonic imaging system vs. Known volume required to fill balloons and excised hearts was evaluated on Volume measurement accuracy (correlation coefficient 0.992). A multiplanar transesophageal ultrasonic imaging system accurately measured the volume of water-filled balloons and excised hearts in vitro.
Synapse has enriched 10 related papers. Consider them for comparative context: