Key result
Speckle tracking echocardiography was feasible in 84.3% of fetuses across three frame rate settings; longitudinal strain increased at higher frame rates while mechanical synchrony remained robust.
Why the study?
The influence of frame rate settings on longitudinal strain and mechanical synchrony values in speckle tracking echocardiography of healthy fetuses was unknown.
Does frame rate setting affect longitudinal strain and mechanical synchrony values in fetal speckle tracking echocardiography?
Observational (n=121)
Does frame rate setting affect longitudinal strain and mechanical synchrony values in fetal speckle tracking echocardiography?
Mechanical synchrony is a robust parameter for fetal speckle tracking echocardiography that is less affected by frame rate, whereas longitudinal strain measurements are significantly influenced by frame rate settings.
Mechanical synchrony may be preferred over strain in fetal STE due to frame rate robustness; leaves open protocol standardization.
Objectives To assess the influence of frame rate settings on longitudinal strain (LS) and mechanical synchrony (SYN) values in Speckle Tracking Echocardiography (STE) of healthy fetuses. Methods In this prospective study, we collected transversal or apical four-chamber-views of 121 healthy fetuses between 20 and 38 weeks of gestation using three different frame rate (FR) settings (≥ 110, 100 ± 10, 60 ± 10 frames per second). We assessed the segmental and the global LS of both ventricles (2C) and of the left ventricle (LV) offline with QLab 10.8 (Philips Medical Systems, Andover, MA, USA ) . Inter- and intraventricular SYN were calculated as time difference in peak myocardial strain between the mid-segments of left and right ventricle (interventricular, 2C_Syn) and lateral wall and septum of the left ventricle (intraventricular, LV_Syn), respectively. Results In 84.3% STE was feasible at all three FR settings. The LS increased in both views at higher FRs to a statistically noticeable extent. SYN measurements and the absolute differences at patient level between the FR settings showed no statistically noticeable alterations. Conclusions STE is feasible at low and high FR settings. SYN emerges to be a robust parameter for fetal STE as it is less affected by the FR. High FRs enable high temporal resolutions and thus an accurate examination of fetal hearts. Future research for the technical implementation of tailored fetal STE software is necessary for reliable clinical application.
No takes yet. Share an insight, caveat, or question.
Haeger et al. (2021) conducted an observational in Healthy fetuses (n=121). Different frame rate settings (≥ 110, 100 ± 10, 60 ± 10 frames per second) was evaluated on Longitudinal strain (LS) and mechanical synchrony (SYN). Speckle tracking echocardiography was feasible in 84.3% of fetuses across three frame rate settings; longitudinal strain increased at higher frame rates while mechanical synchrony remained robust.