Key result
ICD oversensing alerts preempt inappropriate ventricular fibrillation detection far more often than impedance alerts.
Why the study?
How pacing impedance and measures of oversensing change as implantable cardioverter-defibrillator lead fractures progress was unknown.
Do oversensing alerts provide earlier warning of ICD lead pace-sense conductor fracture compared to standard impedance alerts?
Do oversensing alerts provide earlier warning of ICD lead pace-sense conductor fracture compared to standard impedance alerts?
Oversensing alerts provide significantly more advanced warning of impending lead failure than standard impedance alerts because early conductor fracture causes microscopic make-break potentials detectable by sensing algorithms but too transient to trigger conventional impedance thresholds.
Answers are grounded in this paper and cite their sources.
Oversensing alerts may enable earlier ICD lead fracture detection than impedance monitoring; hypothesis-generating in animal models and requires human validation before practice change.
BackgroundPace–sense conductors comprise a pacing coil to the tip electrode and cable to the ring-electrode. Implantable cardioverter-defibrillator (ICD) lead-monitoring diagnostics include pacing impedance (direct current resistance [DCR]) and measures of oversensing. How they change as fractures progress is unknown.ObjectivesTo characterize the relationship between oversensing, impedance, and structural changes in ICD leads developing pace–sense conductor fractures.MethodsWe performed bending tests on 39 leads connected to ICD generators in an electrolyte bath with simulated electrograms. DCR was recorded every 3 minutes; electrograms were telemetered continuously. Twenty-two leads were tested to develop partial or complete fracture criteria confirmed by imaging, using DCR or DCR variability measured by standard deviation (σDCR). Results are reported for 17 other test leads.ResultsInitial oversensing occurred with partial pacing coil fracture vs complete ring cable fracture and correlated with bending-induced DCR peaks. These peaks were too small to be detected by clinical impedance measurements and were characterized by small increases in σDCR (≥0.5 Ω). Impedance threshold alerts occurred at complete pacing coil fracture but only later for ring cable fractures. The oversensing alert triggered before device-detected ventricular fibrillation more frequently than impedance alerts (94% vs 17%; P = .00002).ConclusionsIn conductor fracture, early oversensing corresponds to partial pacing coil fracture or complete ring cable fracture and correlates with transient bending-induced impedance increases, which are detected by impedance variability but too small to trigger clinical impedance alerts. This explains why clinical oversensing alerts provide more warning for device-detected ventricular fibrillation than impedance alerts and suggests how to improve impedance diagnostics based on short-term variability. Pace–sense conductors comprise a pacing coil to the tip electrode and cable to the ring-electrode. Implantable cardioverter-defibrillator (ICD) lead-monitoring diagnostics include pacing impedance (direct current resistance [DCR]) and measures of oversensing. How they change as fractures progress is unknown. To characterize the relationship between oversensing, impedance, and structural changes in ICD leads developing pace–sense conductor fractures. We performed bending tests on 39 leads connected to ICD generators in an electrolyte bath with simulated electrograms. DCR was recorded every 3 minutes; electrograms were telemetered continuously. Twenty-two leads were tested to develop partial or complete fracture criteria confirmed by imaging, using DCR or DCR variability measured by standard deviation (σDCR). Results are reported for 17 other test leads. Initial oversensing occurred with partial pacing coil fracture vs complete ring cable fracture and correlated with bending-induced DCR peaks. These peaks were too small to be detected by clinical impedance measurements and were characterized by small increases in σDCR (≥0.5 Ω). Impedance threshold alerts occurred at complete pacing coil fracture but only later for ring cable fractures. The oversensing alert triggered before device-detected ventricular fibrillation more frequently than impedance alerts (94% vs 17%; P = .00002). In conductor fracture, early oversensing corresponds to partial pacing coil fracture or complete ring cable fracture and correlates with transient bending-induced impedance increases, which are detected by impedance variability but too small to trigger clinical impedance alerts. This explains why clinical oversensing alerts provide more warning for device-detected ventricular fibrillation than impedance alerts and suggests how to improve impedance diagnostics based on short-term variability.
No takes yet. Share an insight, caveat, or question.
Himes et al. (2022) studied developing implantable cardioverter-defibrillator (ICD) lead pace-sense conductor fracture (n=39). Oversensing diagnostic alerts vs. Clinical impedance alerts was evaluated on Alert triggered before inappropriate device-detected ventricular fibrillation (p=0.00002). In developing ICD conductor fractures, clinical oversensing alerts triggered before device-detected ventricular fibrillation more often than standard impedance alerts.