Key result
Direct visualization using a fiberoptic endocardial visualization catheter identified a Thebesian valve in 54% of 98 patients undergoing biventricular device implantation.
Why the study?
Does direct visualization using a fiberoptic endocardial visualization catheter characterize coronary sinus ostium anatomy and facilitate cannulation during biventricular pacemaker implantation?
Observational (n=100)
Does direct visualization using a fiberoptic endocardial visualization catheter characterize coronary sinus ostium anatomy and facilitate cannulation during biventricular pacemaker implantation?
Direct visualization using a fiberoptic catheter reveals that over half of patients have Thebesian valves, which can be successfully traversed for biventricular pacing even when covering most of the ostium.
Answers are grounded in this paper and cite their sources.
Endoscopic CSO visualization reveals valve morphologies that may impede cannulation; leaves open its role in improving implant success.
BACKGROUND: The precise reasons for failure to cannulate the coronary sinus during biventricular device implantation are unknown. Visualization of the coronary sinus ostium during electrophysiology procedures may enhance understanding of how unusual anatomy can affect successful cannulation of the coronary sinus. OBJECTIVES: The aim of this study was to describe the morphology of valves at the coronary sinus ostium (CSO) visualized directly with an illuminated fiberoptic endoscope during implantation of biventricular devices. METHODS: The coronary sinus anatomy of one hundred consecutive patients undergoing implantation of biventricular devices was investigated using a fiberoptic endocardial visualization catheter (EVC). RESULTS: The CSO was clearly visualized in 98 patients using the EVC. A Thebesian valve was seen in 54% of these. Almost all Thebesian valves were positioned at the inferior (61%) or posterior (33%) aspect of the CSO. Only six patients had Thebesian valves that covered more than 70% of the CSO, but all were successfully implanted with a transvenous left ventricular pacing lead after cannulating the coronary sinus under direct visualization. CONCLUSIONS: Over half of patients undergoing biventricular device implantation have identifiable Thebesian valves. Even valves covering the majority of the ostial area may be traversed using direct visualization and modern catheterization techniques.
No takes yet. Share an insight, caveat, or question.
Anh et al. (2007) conducted an observational in Patients undergoing implantation of biventricular devices (n=100). Fiberoptic endocardial visualization catheter (EVC) was evaluated on Presence of a Thebesian valve at the coronary sinus ostium. Direct visualization using a fiberoptic endocardial visualization catheter identified a Thebesian valve in 54% of 98 patients undergoing biventricular device implantation.