Key result
Three-dimensional ultrasound measurement of plaque volume using manual planimetry demonstrated a mean accuracy of 3.1+/-0.9% and intraobserver variability of 4.0+/-1.0% in vascular phantoms.
Why the study?
Does three-dimensional ultrasound accurately and reliably measure carotid plaque volume in test phantoms compared to water displacement?
Does three-dimensional ultrasound accurately and reliably measure carotid plaque volume in test phantoms compared to water displacement?
Three-dimensional ultrasound using serial 2D manual planimetry provides highly accurate and reproducible measurements of carotid plaque volume in phantom models.
No takes yet. Share an insight, caveat, or question.
Phantom validation supports method refinement; leaves open clinical translation to carotid plaque volumetry.
Landry et al. (2002) studied Carotid plaque volume measurement (phantom study). Three-dimensional ultrasound with manual planimetry vs. Water displacement (true volume) was evaluated on Accuracy and variability of plaque volume measurement. Three-dimensional ultrasound measurement of plaque volume using manual planimetry demonstrated a mean accuracy of 3.1+/-0.9% and intraobserver variability of 4.0+/-1.0% in vascular phantoms.