Key result
Segmented MR velocity mapping sequences accurately measured average flow in small moving vessels with a mean error of -2% ± 3% under all simulated cardiac motion conditions.
Why the study?
Does segmented MR velocity mapping accurately measure pulsatile flow in small moving vessels simulating cardiac motion?
Does segmented MR velocity mapping accurately measure pulsatile flow in small moving vessels simulating cardiac motion?
Segmented MR velocity mapping sequences are accurate in measuring average flow in small moving vessels despite motion-induced blurring.
Answers are grounded in this paper and cite their sources.
Supports technical feasibility in motion models; leaves open human validation before clinical use.
The purpose of this study was to investigate the accuracy of conventional, segmented, and echo-shared MR velocity mapping sequences to measure pulsatile flow in small moving vessels using a phantom with simulated cardiac motion. The phantom moved either cyclically in-plane, through-plane, in- and through-plane, or was stationary. The mean error in average flow was -2% +/- 3% (mean +/- SD) for all sequences under all conditions, with or without background correction, as long as the region of interest (ROI) size was equal to the vessel cross-sectional size. Overestimation of flow as a result of an oversized ROI was less than 20%, and independent of field of view (FOV) and matrix, as long as the offset in angle between the imaging plane and flow direction was less than 10 degrees. Segmented velocity mapping sequences are surprisingly accurate in measuring average flow and render flow profiles in small moving vessels despite the blurring in the images due to vessel motion. J. Magn. Reson. Imaging 2001;13:722-728.
No takes yet. Share an insight, caveat, or question.
Arheden et al. (2001) studied this question. Segmented MR velocity mapping sequences vs. Conventional and echo-shared MR velocity mapping sequences was evaluated on Mean error in average flow. Segmented MR velocity mapping sequences accurately measured average flow in small moving vessels with a mean error of -2% ± 3% under all simulated cardiac motion conditions.
Synapse has enriched 10 related papers. Consider them for comparative context: