Key result
Simplified single-beat estimation of the left ventricular end-systolic pressure-volume relationship showed high accuracy compared to the gold-standard method, with an RMSE of 1.7 ± 1.0 mmHg.
Why the study?
Does single-beat estimation accurately reflect the left ventricular end-systolic pressure-volume relationship compared to the multiple-beat vena cava occlusion method in patients with and without severe heart failure?
Cross-Sectional (n=31)
Does single-beat estimation accurately reflect the left ventricular end-systolic pressure-volume relationship compared to the multiple-beat vena cava occlusion method in patients with and without severe heart failure?
Effect estimate: RMSE 1.7 ± 1.0 mmHg
The single-beat method provides an accurate, less invasive estimation of the left ventricular end-systolic pressure-volume relationship across a wide range of ventricular sizes and conditions.
Single-beat ESPVR shows reasonable group-level accuracy in severe HF; leaves open individual reliability for clinical or research adoption.
AIM: The end-systolic pressure-volume relationship (ESPVR) constructed from multiple pressure-volume (PV) loops acquired during load intervention is an established method to asses left ventricular (LV) contractility. We tested the accuracy of simplified single-beat (SB) ESPVR estimation in patients with severe heart failure. METHODS: Nineteen heart failure patients (NYHA III-IV) scheduled for surgical ventricular restoration and/or restrictive mitral annuloplasty and 12 patients with normal LV function scheduled for coronary artery bypass grafting were included. PV signals were obtained before and after cardiac surgery by pressure-conductance catheters and gradual pre-load reductions by vena cava occlusion (VCO). The SB method was applied to the first beat of the VCO run. Accuracy was quantified by the root-mean-square-error (RMSE) between ESPVR(SB) and gold-standard ESPVR(VCO). In addition, we compared slopes (E(ES)) and intercepts (end-systolic volume at multiple pressure levels (70-100 mmHg: ESV(70)-ESV(100)) of ESPVR(SB) vs. ESPVR(VCO) by Bland-Altman analyses. RESULTS: RMSE was 1.7 + or - 1.0 mmHg and was not significantly different between groups and not dependent on end-diastolic volume, indicating equal, high accuracy over a wide volume range. SB-predicted E(ES) had a bias of -0.39 mmHg mL(-1) and limits of agreement (LoA) -2.0 to +1.2 mmHg mL(-1). SB-predicted ESVs at each pressure level showed small bias (range: -10.8 to +9.4 mL) and narrow LoA. Two-way anova indicated that differences between groups were not dependent on the method. CONCLUSION: Our findings, obtained in hearts spanning a wide range of sizes and conditions, support the use of the SB method. This method ultimately facilitates less invasive ESPVR estimation, particularly when coupled with emerging noninvasive techniques to measure LV pressures and volumes.
No takes yet. Share an insight, caveat, or question.
Brinke et al. (2009) conducted a cross-sectional in Severe heart failure (n=31). Simplified single-beat (SB) ESPVR estimation vs. Gold-standard ESPVR(VCO) (multiple pressure-volume loops by vena cava occlusion) was evaluated on Accuracy quantified by the root-mean-square-error (RMSE) between ESPVR(SB) and gold-standard ESPVR(VCO) (RMSE 1.7 ± 1.0 mmHg). Simplified single-beat estimation of the left ventricular end-systolic pressure-volume relationship showed high accuracy compared to the gold-standard method, with an RMSE of 1.7 ± 1.0 mmHg.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: