Key result
Ventricular assistance with the Impella Recover LD blood pump provided adequate flow, but the patient died 48 hours later due to severe lower limb ischemia, with necropsy revealing a subendocardial hematoma.
Case Report (n=1)
No
Answers are grounded in this paper and cite their sources.
This case highlights the potential for endocardial injury, specifically subendocardial hematoma, associated with the use of intracardiac microaxial pumps.
We report the case of a patient unweanable from cardiopulmonary bypass who underwent transaortic implantation of an Impella Recover LD blood pump. The ventricular assistance afforded an adequate flow but the patient died 48 hours after due to severe ischemia of the lower limbs. Chronic atherosclerotic occlusion of the abdominal aorta could be observed during the necropsy study as well as a subendocardial hematoma of the left ventricular wall. Some aspects concerning the pathologic finding observed in the heart are discussed.
No takes yet. Share an insight, caveat, or question.
Fernández et al. (2009) conducted a case report in Postcardiotomy shock (n=1). Impella Recover LD blood pump was evaluated on Clinical outcome. Ventricular assistance with the Impella Recover LD blood pump provided adequate flow, but the patient died 48 hours later due to severe lower limb ischemia, with necropsy revealing a subendocardial hematoma.