Key result
Single lead VDDR pacing provided a stable pacing mode in 97% of patients with complete atrioventricular block over 12 months, with minimal and asymptomatic subclinical atrial undersensing.
Why the study?
Does single lead VDDR pacing provide stable P wave sensing over long-term follow-up in patients with complete atrioventricular block?
Observational (n=30)
Does single lead VDDR pacing provide stable P wave sensing over long-term follow-up in patients with complete atrioventricular block?
Single lead VDDR pacing provides stable long-term P wave sensing in the vast majority of patients with complete AV block, with minimal clinically significant undersensing.
Answers are grounded in this paper and cite their sources.
P-wave sensing stability with single-pass VDDR leads remains unproven; leaves open need for larger studies before routine use in complete AV block.
Optimal function of a single lead P wave synchronous rate adaptive ventricular pacing system (VDDR) requires reliable P wave sensing over time and during daily activities. The stability of P wave sensing and the incidence of sensitivity reprogramming in a single pass lead with a diagonally arranged bipole was assessed in 30 patients with complete atrioventricular block over a follow-up period of 12 +/- 1 months (range 6 months to 3 years). Atrial sensing was assessed during clinic visits, by physical maneuvers (postural changes, breathing, Valsalva maneuver, walking and isometric exercise), maximum treadmill exercise and Holter recordings. P wave amplitude at implantation was 1.21 +/- 0.09 (0.5-3.6) mV, and the atrial sensing threshold remained stable over the entire period of follow-up. Using an atrial sensitivity based on twice the sensing threshold at 1 month, P wave undersensing was found in 2, 4, 3, and 7 patients during clinic visit, physical maneuvers, exercise, and Holter recordings, respectively. Atrial sensitivity reprogramming was performed in three patients based on the correction of undersensing during physical maneuvers. Although eight patients had atrial undersensing on Holter recordings, the number of undersensed P waves was small (total 101 beats or 0.013% +/- 0.001% of total ventricular beats) and no patient was symptomatic. One patient had intermittent atrial undersensing at the highest sensitivity, but the VDDR mode was still functional most of the time. No patient had myopotential interference at the programmed sensitivity. One patient developed chronic atrial fibrillation and was programmed to the VVIR mode. Thus, single lead VDDR pacing is a stable pacing mode in 97% of patients.(ABSTRACT TRUNCATED AT 250 WORDS)
No takes yet. Share an insight, caveat, or question.
Lau et al. (1994) conducted an observational in Complete atrioventricular block (n=30). Single lead VDDR pacing was evaluated on Stability of P wave sensing and incidence of sensitivity reprogramming. Single lead VDDR pacing provided a stable pacing mode in 97% of patients with complete atrioventricular block over 12 months, with minimal and asymptomatic subclinical atrial undersensing.