Key result
Intracardiac shocks of 360 joules delivered via a temporary pacing catheter successfully defibrillated a patient with intractable ventricular fibrillation refractory to transthoracic shocks.
Why the study?
Does intracardiac shock via a temporary pacing catheter convert intractable ventricular fibrillation in a patient refractory to transthoracic defibrillation?
Case Report (n=1)
Does intracardiac shock via a temporary pacing catheter convert intractable ventricular fibrillation in a patient refractory to transthoracic defibrillation?
Intracardiac shocks using a standard temporary pacing catheter can successfully convert ventricular fibrillation that is refractory to standard transthoracic defibrillation.
Direct application of electrical current to the heart utilizing special equipment such as implantable defibrillators or specially designed catheters has been useful in patients with recurrent ventricular tachyarrhythmias. We describe a patient who developed intractable ventricular fibrillation (VF) during the course of electrophysiologic testing. VF continued for 50 minutes, during which time multiple transthoracic shocks of 360 joules failed to convert the arrhythmia. Intracardiac shocks were then delivered via a transvenous 6 French standard quadripolar pacing catheter positioned in the right ventricular apex by direct application of the defibrillator paddle to the catheter, with the patient lying on the posterior paddle. Energy of up to 300 joules failed to defibrillate the patient, but use of 360 joules delivered in this fashion defibrillated the patient twice. Despite the protracted time of cardiac arrest, the patient recovered without sequelae. We report this case to demonstrate that VF refractory to transthoracic defibrillation may be converted by intracardiac shocks using a temporary pacing catheter without special equipment.
No takes yet. Share an insight, caveat, or question.
Mann et al. (1985) conducted a case report in Intractable ventricular fibrillation (n=1). Intracardiac shocks via a temporary pacing catheter vs. Transthoracic shocks was evaluated on Defibrillation of ventricular fibrillation. Intracardiac shocks of 360 joules delivered via a temporary pacing catheter successfully defibrillated a patient with intractable ventricular fibrillation refractory to transthoracic shocks.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: