Why the study?
Does a custom built disposable catheter for internal cardioversion and defibrillation successfully convert VT or VF in patients?
Does a custom built disposable catheter for internal cardioversion and defibrillation successfully convert VT or VF in patients?
A custom-built disposable catheter for internal cardioversion and defibrillation successfully converted >95% of VT/VF episodes with low energy, suggesting a practical method for tachyarrhythmia management.
May support low-energy internal defibrillation via disposable catheter; hypothesis-generating and requires prospective trials before clinical adoption.
The use of transvenous ICD systems and the recent advances in atrial defibrillation techniques have heightened interest in internal defibrillation. However, most shocks for induced or spontaneous arrhythmias in patients without devices are still delivered transthoracically using high energy. We describe the history of temporary internal defibrillation techniques and report the initial clinical results with a custom built disposable catheter for internal cardioversion and defibrillation. This prototype successfully converted more than 95% of 109 episodes of VT or VF in 28 patients, with biphasic energies < or = 20 J. A newer disposable catheter, using 40-wire Matrix technology as the defibrillating electrode, has design features that provide high surface area, low impedance, and low current density when compared to other leads used for similar purposes. Temporary internal cardioversion-defibrillation of induced and spontaneous arrhythmias using such catheter designs is likely to be widely applicable to patients undergoing electrophysiology procedures and to those in critical care units prone to tachyarrhythmias.
No takes yet. Share an insight, caveat, or question.
Luceri et al. (1997) studied this question.