Key result
Immediately following rapid preload changes, wearable carotid Doppler ultrasound and non-invasive pulse contour analysis detected hemodynamic changes with high accuracy (AUROC 0.98), whereas bioreactance had lower initial accuracy (AUROC 0.75) that improved over time.
Why the study?
The study was conducted to describe the temporal concordance of 3 non-invasive hemodynamic monitors during rapid preload changes.
How do pulse contour analysis, bioreactance, and carotid Doppler compare in their temporal concordance and accuracy for detecting rapid preload changes in healthy volunteers?
Observational (n=7)
No
How do pulse contour analysis, bioreactance, and carotid Doppler compare in their temporal concordance and accuracy for detecting rapid preload changes in healthy volunteers?
Absolute Event Rate: 0.98% vs 0.75%
Carotid Doppler ultrasound and non-invasive pulse contour detect rapid hemodynamic changes immediately with high accuracy, whereas bioreactance exhibits an algorithm lag.
PURPOSE: We describe the temporal concordance of 3 hemodynamic monitors. MATERIALS AND METHODS: Healthy volunteers performed preload changes while simultaneously wearing a non-invasive, pulse-contour stroke volume (SV) monitor, a bioreactance SV monitor and a wireless, wearable Doppler ultrasound patch over the common carotid artery. The sensitivity and specificity for detecting preload change over 3 temporal windows (early, middle and late) was assessed. RESULTS: 40 preload changes were recorded in total (20 increase, 20 decrease). Immediately, the wearable Doppler had high sensitivity (100%) and specificity (100%) for detecting preload change with an area under the receiver operator curve (AUROC) of 0.98 for both velocity time integral (VTI, 10.5% threshold) and corrected flow time (FTc, 2.5% threshold). The sensitivity, specificity and AUROC for non-invasive pulse contour were equally good (9% SV threshold). For bioreactance, a 13% SV threshold immediately detected preload change with a sensitivity, specificity and AUROC of 60%, 95% and 0.75, respectively. After two SV outputs following preload change, the sensitivity, specificity and AUROC of bioreactance improved to 70%, 90% and 0.85, respectively. CONCLUSIONS: Carotid Doppler ultrasound and non-invasive pulse contour detected rapid hemodynamic change with equal accuracy; bioreactance improved over time. Algorithm-lag should be considered when interpreting clinical studies.
No takes yet. Share an insight, caveat, or question.
Kenny et al. (2022) conducted an observational in Healthy volunteers (n=7). Wearable carotid Doppler ultrasound and non-invasive pulse contour analysis vs. Bioreactance was evaluated on Area under the receiver operator curve (AUROC) for detecting preload change in the early temporal window. Immediately following rapid preload changes, wearable carotid Doppler ultrasound and non-invasive pulse contour analysis detected hemodynamic changes with high accuracy (AUROC 0.98), whereas bioreactance had lower initial accuracy (AUROC 0.75) that improved over time.