Key result
Automated mitral chordal ePTFE suturing in human heart specimens required a mean of 3.6 ± 1.5 minutes per chordal replacement, with good control of suture delivery in 95.8% of leaflet sutures.
Why the study?
The study evaluated the ergonomics and time requirements of a novel automated suturing and titanium fastener deployment technology for chordal replacement in human heart specimens using open and MICS simulators.
Does a novel automated suturing technology provide efficient and accurate mitral chordal replacement in human heart specimens?
Does a novel automated suturing technology provide efficient and accurate mitral chordal replacement in human heart specimens?
A novel automated suturing technology for mitral chordal replacement demonstrated accurate placement and efficient handling in human cadaveric heart models, supporting its potential for minimally invasive mitral valve repair.
Supports preclinical feasibility in specimens; leaves open translation to clinical mitral valve repair.
Objective: This study evaluated the ergonomics and time requirements of using a novel automated suturing and titanium fastener deployment technology for chordal replacement in human heart specimens in open and minimally invasive cardiac surgery (MICS) simulators. Methods: Five cardiac surgeons used novel, manually powered expanded polytetrafluoroethylene (ePTFE) suturing devices to automate suture placement between mitral leaflets and papillary muscles in explanted cadaver hearts, along with customized titanium fastener delivery devices to secure suture and trim suture tails. This mitral chordal replacement test was conducted using surgical models simulating open and MICS mitral repair access. The study was approved by the institutional ethical board. Results: After a brief introduction to this technique using plastic models, study surgeons performed 48 chordal replacements in human mitral valves, placing 18 in an open model and 30 in a right minithoracotomy model. The time range to complete a single chordal replacement was between 55 s and 8 min, with an overall mean duration of 3.6 ± 1.5 min. No difference in duration of implantation was recorded for the MICS and open sternotomy simulators used. Good control of suture delivery was reported in 95.8% ( n = 46) of leaflet aspect of the sutures and in 100% ( N = 48) of papillary muscle sutures. Conclusions: Automated mitral chordal ePTFE suturing simulated through open and MICS access demonstrated quality handling and accurate placement of sutures in human heart specimens. A clinical trial using this technology is currently ongoing. This innovation may present an important advance facilitating enhanced minimally invasive mitral valve repair.
No takes yet. Share an insight, caveat, or question.
Werner et al. (2022) studied Mitral chordal replacement (preclinical model) (n=5). Automated suturing and titanium fastener deployment technology was evaluated on Ergonomics and time requirements for chordal replacement. Automated mitral chordal ePTFE suturing in human heart specimens required a mean of 3.6 ± 1.5 minutes per chordal replacement, with good control of suture delivery in 95.8% of leaflet sutures.