Key result
Active pectoral defibrillation thresholds remained stable over the first 2 months postimplantation, decreasing from 9.9+/-5.5 J at implant to 7.6+/-5.5 J at follow-up.
Why the study?
Are defibrillation thresholds stable over time with an active pectoral lead system in patients receiving an ICD?
Observational (n=46)
Are defibrillation thresholds stable over time with an active pectoral lead system in patients receiving an ICD?
Answers are grounded in this paper and cite their sources.
Active pectoral defibrillation thresholds are stable over the first 2 months postimplantation, questioning the need for routine serial defibrillation threshold testing.
INTRODUCTION: Monophasic defibrillation thresholds rise over time with a variety of lead systems. These chronic changes are attenuated or eliminated by biphasic waveforms, although the effect appears dependent upon the lead system. With the downsizing of pulse generator size to allow for routine pectoral implantation, active can lead systems have now become standard. However, the temporal stability of such lead systems has not been evaluated previously. METHODS AND RESULTS: This study was a prospective assessment of the changes of active pectoral defibrillation thresholds over time. Thresholds were measured at implant, predischarge, and at a mean follow-up of 50 days in 46 patients with a uniform testing protocol and shock polarity. The lead system was a dual-coil Endotak DSP lead with an active pectoral pulse generator. Defibrillation thresholds were 9.9+/-5.5 J at implantation, 8.5+/-6.0 J predischarge, and 7.6+/-5.5 J at follow-up (ANOVA, P = 0.007). Moreover, only two patients developed an increased threshold > 5 J, and no patient had an inadequate safety margin at follow-up. CONCLUSION: These results indicate that active pectoral defibrillation thresholds are stable over the first 2 months postimplantation and question the need for routine serial defibrillation threshold testing.
No takes yet. Share an insight, caveat, or question.
Olsovsky et al. (1998) conducted an observational in Patients receiving an active pectoral lead system (n=46). Active pectoral lead system vs. Baseline (implantation) was evaluated on Defibrillation thresholds (p=0.007). Active pectoral defibrillation thresholds remained stable over the first 2 months postimplantation, decreasing from 9.9+/-5.5 J at implant to 7.6+/-5.5 J at follow-up.