Does sequential pulse defibrillation threshold change with maturation of the electrode tissue interface up to 12 weeks after implantation?
Does sequential pulse defibrillation threshold change with maturation of the electrode tissue interface up to 12 weeks after implantation?
Sequential pulse defibrillation thresholds remain stable for up to 12 weeks post-implantation, suggesting that testing at implant is predictive of future device performance.
Answers are grounded in this paper and cite their sources.
Implant testing may predict chronic thresholds; leaves open generalizability to modern leads and randomized confirmation.
We have shown that sequential pulse defibrillation threshold voltage and total delivered energy do not change with maturation of the electrode tissue interface for up to 12 weeks after implantation of two different electrode configurations. This result is important to predict the future performance of an implantable defibrillator that is tested only at implant.
No takes yet. Share an insight, caveat, or question.
Kallok et al. (1986) studied this question.