Key result
Late insulation breaks in endocardial sensing leads for cardioverter defibrillators caused inappropriate discharges 10-30 months post-implantation in four patients.
Case Report (n=4)
Answers are grounded in this paper and cite their sources.
Late insulation breaks in endocardial sensing leads can cause inappropriate discharges 10-30 months post-implantation and should be considered in the differential diagnosis of device failure.
Despite their similarity to permanent pacemaker leads, endocardial sensing leads for cardioverter defibrillators have a relatively high failure rate. We describe four patients with endocardial rate sensing leads who developed inappropriate discharges 10-30 months after implantation due to small breaks in the lead insulation. This problem may become increasingly common as the number of cardioverter defibrillator implants with transvenous leads continues to grow and should be considered in the differential diagnosis of late sensing failure or inappropriate device discharges.
No takes yet. Share an insight, caveat, or question.
Peters et al. (1995) conducted a case report in Patients with endocardial rate sensing leads for cardioverter defibrillators (n=4). Late insulation break in endocardial sensing leads was evaluated on Inappropriate discharges. Late insulation breaks in endocardial sensing leads for cardioverter defibrillators caused inappropriate discharges 10-30 months post-implantation in four patients.