Why the study?
Does ICD sensing lead repair prevent new lead abnormalities as effectively as lead replacement in patients with sensing lead insulation breaks?
Population
52 patients undergoing implantable cardioverter defibrillator sensing lead repair or replacement for…
Comparison
ICD sensing lead repair utilizing a commercially… vs ICD sensing lead replacement with new sensing…
Design
Cohort
Follow-up
44 +/- 23 months
Authors
Loading...
Lead repair may be a viable alternative when feasible; leaves open need for randomized trials to confirm equivalence.
Does ICD sensing lead repair prevent new lead abnormalities as effectively as lead replacement in patients with sensing lead insulation breaks?
Repairing ICD sensing leads with insulation breaks provides equivalent long-term durability compared to lead replacement, suggesting repair is a viable strategy when technically feasible.
Mahapatra et al. (2003) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: