Key result
A single-coil active fix implantable defibrillator lead model exhibited late malfunction in two patients due to conductor separation visible on chest radiographs.
Highlights a specific mechanism of late failure (conductor separation) in a single-coil active fix implantable defibrillator lead.
Answers are grounded in this paper and cite their sources.
Alerts clinicians to inspect radiographs for conductor separation with this lead; leaves open population-level failure rates.
Two patients with the same model of single-coil active fix implantable defibrillator lead presented with evidence of lead malfunction, in one case with an abrupt rise in pacing impedance and threshold and in the other with evidence of over-sensing. In both cases, the chest radiograph showed separation of conductors in the defibrillation lead from the main body of the lead.
No takes yet. Share an insight, caveat, or question.
Richards et al. (2010) studied this question. A single-coil active fix implantable defibrillator lead model exhibited late malfunction in two patients due to conductor separation visible on chest radiographs.