Key result
Abdominal ultrasonography produced highly reliable measurements of subcutaneous and visceral adipose tissue at 12 weeks gestation, with inter-observer limits of agreement within 0.5 cm.
Why the study?
Excess abdominal adiposity causes metabolic disturbances in pregnancy, but accurate measurement methods are limited due to procedure risks.
Is ultrasound a reliable method for measuring subcutaneous and visceral adipose tissue thickness in pregnant women at 12 weeks gestation?
Cross-Sectional (n=30)
No
Is ultrasound a reliable method for measuring subcutaneous and visceral adipose tissue thickness in pregnant women at 12 weeks gestation?
Ultrasound provides reliable and repeatable measurements of subcutaneous and visceral adipose tissue thickness during early pregnancy.
Supports reliable ultrasound adipose assessment at 12 weeks; leaves open outcome validation before clinical adoption.
BACKGROUND: Excess abdominal adiposity cause metabolic disturbances, particularly in pregnancy. Methods of accurate measurement are limited in pregnancy due to risks associated with these procedures. This study outlines a non-invasive methodology for the measurement of adipose tissue in pregnancy and determines the intra- and inter-observer reliability of ultrasound (US) measurements of the two components of adipose tissue (subcutaneous (SAT) and visceral adipose tissue (VAT)) within a pregnant population. METHODS: Thirty pregnant women were recruited at the end of their first trimester, from routine antenatal clinic at the University Maternity Hospital Limerick, Ireland. Measurements of adipose tissue thickness were obtained using a GE Voluson E8 employing a 1-5 MHz curvilinear array transducer. Two observers, employing methodological rigour in US technique, measured thickness of adipose tissue three times, and segmented the US image systematically in order to define measurements of SAT and VAT using specifically pre-defined anatomical landmarks. RESULTS: Intra-observer and inter-observer precision was assessed using Coefficient of Variation (CV). Measurements of SAT and total adipose for both observers were < 5% CV and < 10% CV for VAT in measures by both observers. Inter-observer reliability was assessed by Limits of Agreement (LoA). LoA were determined to be - 0.45 to 0.46 cm for SAT and - 0.34 to 0.53 cm for VAT values. Systematic bias of SAT measurement was 0.01 cm and 0.10 cm for VAT. Inter-observer precision was also assessed by coefficient of variation (CV: SAT, 3.1%; VAT, 7.2%; Total adipose, 3.0%). CONCLUSION: Intra-observer precision was found to be acceptable for measures of SAT, VAT and total adipose according to anthropometric criterion, with higher precision reported in SAT values than in VAT. Inter-observer reliability assessed by Limits-Of-Agreement (LoA) confirm anthropometrically reliable to 0.5 cm. Systematic bias was minimal for both measures, falling within 95% confidence intervals. These results suggest that US can produce reliable, repeatable and accurate measures of SAT and VAT during pregnancy.
No takes yet. Share an insight, caveat, or question.
Cremona et al. (2019) conducted a cross-sectional in Pregnancy (n=30). Abdominal ultrasonography vs. Inter-observer and intra-observer comparison was evaluated on Intra- and inter-observer reliability of ultrasound measurements of subcutaneous and visceral adipose tissue. Abdominal ultrasonography produced highly reliable measurements of subcutaneous and visceral adipose tissue at 12 weeks gestation, with inter-observer limits of agreement within 0.5 cm.