Key result
The proposed matched filtering method reduced the average error of Doppler radar-based heart rate measurement from 46.21% to 0.96% across different subjects and distances.
Why the study?
Vital sign radar detection accuracy of heartbeat rate is heavily affected by respiration frequency harmonics.
Does a waveform-driven matched filtering method improve heart rate detection accuracy in a 10-GHz non-contact CW Doppler radar system compared to standard processing?
Does a waveform-driven matched filtering method improve heart rate detection accuracy in a 10-GHz non-contact CW Doppler radar system compared to standard processing?
Absolute Event Rate: 0.96% vs 46.21%
A novel matched filtering method based on polynomial fitting significantly improves the accuracy of non-contact Doppler radar heart rate measurements by removing respiration harmonic interference.
May improve non-contact radar HR accuracy; leaves open clinical validation before practice adoption.
To enhance the vital sign radar’s detection accuracy of heartbeat rate (HR), which is heavily affected by the harmonics of respiration frequency, this paper proposes a waveform-driven matched filtering method based on polynomial fitting extraction. The merit of this approach lies in that, the quasi-ideal matching impulse approximating the actual heartbeat signal can be readily retrieved by subtracting the polynomial fitted waveform from a received signal, which provides high adaptability for individual subjects. Essentially, this extraction greatly removes the harmonic interference originated from respiration, thus considerably improving the HR detection accuracy. Simulations are performed to acquire the proper fitting order and the effective impulse duration. Guided by these two parameters, a 10-GHz non-contact continuous-wave (CW) Doppler radar system with typical microwave laboratory instruments is constructed. Experimental results confirmed the effectiveness of the proposed method, showing that the average errors of HR can be reduced from 46.21% to 0.96% under different subjects and distances, and from 38.05% to 0.90% in continuous measurement for one subject.
No takes yet. Share an insight, caveat, or question.
Yang et al. (2020) conducted a letter in Heart rate measurement (n=2). Waveform-driven matched filtering method based on polynomial fitting extraction vs. Without matched filter was evaluated on Average error of heart rate measurement. The proposed matched filtering method reduced the average error of Doppler radar-based heart rate measurement from 46.21% to 0.96% across different subjects and distances.