Key result
A chair-based cuffless blood pressure monitoring system estimated systolic and diastolic BP with a mean difference of 0.2 ± 5.8 mmHg and 0.4 ± 5.7 mmHg compared to a reference cuff monitor.
Why the study?
Does a chair-based unobtrusive cuffless blood pressure monitoring system accurately estimate blood pressure compared to a standard cuff-type monitor?
Does a chair-based unobtrusive cuffless blood pressure monitoring system accurately estimate blood pressure compared to a standard cuff-type monitor?
A novel chair-based cuffless blood pressure monitoring system using pulse arrival time provides accurate BP estimations comparable to cuff-type monitors, showing potential for long-term home monitoring.
Answers are grounded in this paper and cite their sources.
Offers preliminary accuracy for chair-based cuffless BP monitoring; leaves open larger validation before clinical use.
In this paper, we present an unobtrusive cuffless blood pressure (BP) monitoring system based on pulse arrival time (PAT) for facilitating long-term home BP monitoring. The proposed system consists of an electrocardiograph (ECG), a photoplethysmograph (PPG), and a control circuit with a Bluetooth module, all of which are mounted on a common armchair to measure ECG and PPG signals from users while sitting on the armchair in order to calculate continuous PAT. Considering the good linear correlation of systolic BP (SBP) and the nonlinear correlation of diastolic BP (DBP) with PAT, a new BP estimation method was proposed. Ten subjects underwent BP monitoring experiments involving stationary sitting on a chair, lying on a bed, and pedaling using an ergometer in order to assess the accuracy of the estimated BP. A cuff-type BP monitor was used as reference in the experiments. Results showed that the mean difference of the estimated SBP and DBP was within 0.2 ± 5.8 mmHg ( p < 0.00001) and 0.4 ± 5.7 mmHg ( p < 0.00001), respectively, and the mean absolute difference of the estimated SBP and DBP were 4.4 and 4.6 mmHg, respectively, compared to references. Additionally, five subjects participated in data collections consisting of sitting on a chair twice a day for one month. Compared to the reference, the difference did not obviously increase along with time, even though individualized calibration was executed only once at the beginning. These results suggest that the proposed system has quite the potential for long-term home BP monitoring.
No takes yet. Share an insight, caveat, or question.
Tang et al. (2016) studied Blood pressure monitoring (n=10). Chair-based unobtrusive cuffless blood pressure monitoring system vs. Cuff-type BP monitor was evaluated on Accuracy of estimated systolic and diastolic blood pressure (Mean difference 0.2 ± 5.8 mmHg for SBP and 0.4 ± 5.7 mmHg for DBP, p=<0.00001). A chair-based cuffless blood pressure monitoring system estimated systolic and diastolic BP with a mean difference of 0.2 ± 5.8 mmHg and 0.4 ± 5.7 mmHg compared to a reference cuff monitor.