Key result
The CNAP2GO continuous noninvasive blood pressure measurement device showed a mean difference of -1.0 mmHg compared to invasive arterial catheter measurements.
Why the study?
Current wearable methods analyzing time components lack clinical accuracy, whereas existing technologies for direct continuous blood pressure measurement are too bulky.
Does the CNAP2GO continuous noninvasive blood pressure monitor accurately measure mean blood pressure compared to an invasive arterial catheter in surgical patients?
Observational (n=46)
No
Does the CNAP2GO continuous noninvasive blood pressure monitor accurately measure mean blood pressure compared to an invasive arterial catheter in surgical patients?
The CNAP2GO device provides continuous noninvasive blood pressure monitoring with clinically acceptable accuracy compared to invasive arterial lines, demonstrating feasibility for miniaturized wearable applications.
Answers are grounded in this paper and cite their sources.
May enable continuous wearable BP monitoring; leaves open clinical accuracy and adoption pending validation trials.
Wearable sensors to continuously measure blood pressure and derived cardiovascular variables have the potential to revolutionize patient monitoring. Current wearable methods analyzing time components (e.g., pulse transit time) still lack clinical accuracy, whereas existing technologies for direct blood pressure measurement are too bulky. Here we present an innovative art of continuous noninvasive hemodynamic monitoring (CNAP2GO). It directly measures blood pressure by using a volume control technique and could be used for small wearable sensors integrated in a finger-ring. As a software prototype, CNAP2GO showed excellent blood pressure measurement performance in comparison with invasive reference measurements in 46 patients having surgery. The resulting pulsatile blood pressure signal carries information to derive cardiac output and other hemodynamic variables. We show that CNAP2GO can self-calibrate and be miniaturized for wearable approaches. CNAP2GO potentially constitutes the breakthrough for wearable sensors for blood pressure and flow monitoring in both ambulatory and in-hospital clinical settings.
No takes yet. Share an insight, caveat, or question.
Fortin et al. (2021) conducted an observational in Patients undergoing neurosurgery with general anesthesia (n=46). CNAP2GO continuous noninvasive blood pressure measurement vs. Invasive radial arterial catheter was evaluated on Mean blood pressure (mBP) agreement (MD -1.0 mmHg, 95% CI -14.8 to 12.7). The CNAP2GO continuous noninvasive blood pressure measurement device showed a mean difference of -1.0 mmHg compared to invasive arterial catheter measurements.
Synapse has enriched 10 related papers. Consider them for comparative context: