Key result
Multiplex lateral flow devices LFD1 and LFD2 demonstrated diagnostic sensitivities of 87.5% and 70% respectively, and 100% specificity for detecting foot-and-mouth disease viruses.
Why the study?
Serotype identification is crucial for foot-and-mouth disease transmission evaluation and vaccine selection, but existing lateral flow device prototypes had limited serotyping capacity.
Do multiplex lateral flow devices accurately detect and serotype foot-and-mouth disease viruses compared to standard laboratory methods?
Do multiplex lateral flow devices accurately detect and serotype foot-and-mouth disease viruses compared to standard laboratory methods?
Multiplex lateral flow devices demonstrate high specificity and acceptable sensitivity for rapid field detection and serotyping of foot-and-mouth disease viruses.
Answers are grounded in this paper and cite their sources.
May support rapid FMD serotyping in endemic settings; leaves open larger field validation before routine adoption.
Lateral Flow Devices (LFDs) represent a simple tool for the rapid diagnosis of FMD, particularly in endemic regions, often lacking adequate and equipped laboratories. Other experimental prototypes offer a user-friendly FMD confirmation tool, but serotype identification is crucial for transmission patterns evaluation and for vaccine selection and their serotyping capacity was limited. This study describes the validation of two multiplex devices based on well-characterized monoclonal antibodies: LFD1 (O, A and Asia1) and LFD2 (SAT1 and SAT2). First, the ability of the LFDs to recognize the widest spectrum of strains was assessed. Eighty-three tissue-cultured FMDV strains were tested with LFD1 and 15 with LFD2, confirming the capability to recognize and serotype a broad range of lineages and sub-lineages belonging to all 7 endemic pools. The diagnostic sensitivity related to FMDV detection was evaluated using 224 real-time RT-PCR positive samples collected in endemic countries, resulting in 87.5 % for LFD1 (140/160 detected) and 70 % for LFD2 (45/64 detected). The diagnostic sensitivity of serotyping capabilities was assessed by testing field samples in parallel with an antigen-ELISA kit: applied on 178 samples, LFD1 correctly typed 96 %, 94 % and 83 % of samples positive for type O (n = 95), A (n = 54) and Asia1 (n = 29) respectively, while LFD2 correctly typed 77 % SAT1 (n = 52) and 82 % SAT2 (n = 40) in 92 positive field samples. A 100 % diagnostic specificity was found testing 60 epithelium homogenates collected from animals negative for FMD. Obtained data highlighted the devices' potential value for field use to support the enhanced surveillance of FMD.
No takes yet. Share an insight, caveat, or question.
Foglia et al. (2025) studied Foot-and-mouth disease (n=284). Multiplex lateral flow devices (LFD1 and LFD2) vs. Real-time RT-PCR and antigen-ELISA was evaluated on Diagnostic sensitivity for FMDV detection. Multiplex lateral flow devices LFD1 and LFD2 demonstrated diagnostic sensitivities of 87.5% and 70% respectively, and 100% specificity for detecting foot-and-mouth disease viruses.