Why the study?
Does the addition of a defibrillation electrode in the low right atrium to a right ventricular lead reduce ventricular defibrillation thresholds in patients receiving an ICD?
Does the addition of a defibrillation electrode in the low right atrium to a right ventricular lead reduce ventricular defibrillation thresholds in patients receiving an ICD?
Adding a defibrillation coil in the low right atrium reduces total shock impedance but fails to reduce ventricular defibrillation thresholds in patients receiving an ICD.
Answers are grounded in this paper and cite their sources.
Adding low RA coil fails to lower V-DFT despite reduced impedance; challenges impedance reduction as surrogate for defibrillation efficacy in ICD lead placement.
Transvenous unipolar active can defibrillation systems have proven to be effective in treating ventricular tachyarrhythmias. However, a further reduction of ventricular defibrillation thresholds (V-DFT) would increase the longevity, reduce the size of pulse generators, and help to avoid additional leads in patients with inacceptable high V-DFTs. In a finite difference computer model, the extension of the right ventricular (RV) defibrillation coil into the low right atrium led to a 40% reduction of unipolar V-DFT. To evaluate this finding, we conducted a prospective, randomized study in 11 patients receiving an ICD. Extension of the RV electrode was simulated by adding a second coil placed in the low right atrium with the same polarity. Using a binary search protocol, V-DFT was determined with and without the additional electrode in each patient. Total shock impedance was significantly lower in the two coil (low RA) configuration, compared to the single coil (RV) configuration. Corresponding values were 49.9 +/- 6.7 Ohm and 61.1 +/- 9.3 Ohm, respectively (P < 0.01, paired t-test). However, there was no reduction, but even a nonsignificant increase in V-DFTs. Mean V-DFT in the RV configuration was 12.0 +/- 5.6 J and 16.3 +/- 7.8 J in the low RA configuration (P = 0.09, paired t-test). Despite a reduction in total impedance, the addition of a defibrillation coil in the low right atrium does not reduce ventricular defibrillation thresholds.
No takes yet. Share an insight, caveat, or question.
Rüb et al. (2004) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: