A combined subcutaneous array and azygos vein lead may be an effective strategy for reducing high defibrillation thresholds when standard approaches fail.
May offer bailout in refractory high DFT; case report leaves open systematic validation before practice change.
A 72-year-old man with nonischemic cardiomyopathy was referred because his implantable cardioverter defibrillator had failed to terminate spontaneous ventricular fibrillation (VF). Defibrillation threshold (DFT) testing confirmed that 830-V shocks failed to defibrillate VF despite optimization of the biphasic waveform and reversal of shock polarity. The placement of a new right ventricular lead and the addition of a subcutaneous array failed to defibrillate VF at 830 V. The combination of a subcutaneous array and azygos vein coil successfully defibrillated VF. The mechanism for successful DFT reduction was likely greater current supplied to the posterior basal left ventricle by the azygos vein lead.