Key result
Placement of a single-coil lead into the hemiazygous vein in a patient with a right-sided ICD reduced defibrillation thresholds from >35 J to <25 J.
Why the study?
Does hemiazygous coil placement reduce high defibrillation thresholds in a patient with a right-sided ICD?
Case Report (n=1)
Does hemiazygous coil placement reduce high defibrillation thresholds in a patient with a right-sided ICD?
Hemiazygous coil placement can effectively lower high defibrillation thresholds in patients with right-sided ICDs by improving the shocking vector.
Answers are grounded in this paper and cite their sources.
May lower high DFTs in right-sided ICDs; hypothesis-generating and should not yet change practice.
A 41-year-old man underwent implantation of a right-sided implantable cardioverter defibrillator after removal of an infected left-sided system. Defibrillation threshold (DFT) testing on the right-sided system failed to convert ventricular fibrillation at maximum device output (35 J) compared with a DFT of less than 15 J on the previous left-sided system. A single-coil lead was selectively placed into the hemiazygous vein, which courses leftward of the spine in a posterior-anterior projection, resulting in an improved shocking vector and reduction in DFTs to less than 25 J.
No takes yet. Share an insight, caveat, or question.
Nilsson et al. (2010) conducted a case report in High-defibrillation thresholds in right-sided ICD (n=1). Hemiazygous coil placement vs. Right-sided system alone was evaluated on Defibrillation threshold (DFT). Placement of a single-coil lead into the hemiazygous vein in a patient with a right-sided ICD reduced defibrillation thresholds from >35 J to <25 J.
Synapse has enriched 10 related papers. Consider them for comparative context: