Why the study?
Does implantation of early-generation sirolimus- or paclitaxel-eluting stents result in delayed neointimal growth or late catch-up phenomenon between 3 and 9 years post-PCI in patients with stable CAD?
Does implantation of early-generation sirolimus- or paclitaxel-eluting stents result in delayed neointimal growth or late catch-up phenomenon between 3 and 9 years post-PCI in patients with stable CAD?
Serial OCT assessment at 3 and 9 years post-PCI with early-generation sirolimus- and paclitaxel-eluting stents showed stable neointimal thickness and strut coverage, suggesting an absence of the 'very late catch-up' phenomenon in event-free patients.
Answers are grounded in this paper and cite their sources.
Serial OCT up to 9 years after early-generation DES may clarify delayed healing; leaves open relevance to contemporary platforms.
Early-generation drug-eluting stents (DES) have been demonstrated to delay vascular healing. Limited optical coherence tomography (OCT) data on the very long-term neointimal response after DES implantation are available. The aim of this study was a serial OCT assessment of neointimal thickness, stent strut coverage, malapposition, and protrusion as markers of neointimal response at 3 and 9 years after implantation of sirolimus-eluting stents (SES) and paclitaxel-eluting stents (PES). In this single-centre, longitudinal study consecutive patients undergoing elective PCI with SES or PES were included. OCT analysis was performed after 3 and 9 years by the independent core laboratory. A total of 22 subjects (8 SES and 14 PES) underwent an OCT assessment at 3 and 9 years post index procedure. The lumen, neointimal and malapposition area and the neointimal thickness (SES ∆50 µm, p = 0.195, PES ∆10 µm, p = 0.951) did not change significantly over the 6 year follow-up. No differences in the incidence of uncovered, malapposed or protruding struts were found in each type of stent. At 3 and 9 years after PCI, implantation of early-generation SES and PES may be associated with similar neointimal thickness, strut coverage, malapposition and protrusion, as assessed by serial OCT examination among patients with uneventful follow-up at 3 years post procedure. The small size of the study warrants judicious interpretation of our results and confirmation in larger multimodality imaging studies, including patients treated with contemporary stent platforms.
No takes yet. Share an insight, caveat, or question.
Tomaniak et al. (2018) studied this question.