This editorial highlights that measuring the mid-line (modal velocity) of tissue Doppler tracings is more accurate than the commonly used outer edge, while calling attention to significant machine-related biases.
Tissue Doppler, an echocardiographic modality measuring velocities of solid cardiac structures by pulsed-wave or colour Doppler, after very early experiments1 was introduced clinically in the early 1990s2,3 and is one of the very few recent technical innovations in echocardiography to gain quick and near-ubiquitous clinical implementation. The most widely adopted application is the registration of the mitral annular baso-apical velocity profile, at the basal septal or basal lateral left ventricular wall, which gives quantitative information on both diastolic and systolic global left ventricular function and is now part of the standard echo examination.4 These tissue velocities have also been shown to be useful to identify subclinical cardiomyopathies, such as hypertrophic cardiomyopathy,5,6 amyloidosis,7 or drug cardiotoxicity.8 Recent data show a strong prognostic value after myocardial infarction.9 Another established application is right ventricular free wall velocity as a parameter of right ventricular function. Technically, tissue Doppler differs from standard pulsed-wave Doppler by using a low-pass frequency filter to keep low-velocity signals from tissue and to reject the high-velocity signals from blood. The detected velocities are low–even for right ventricular basal velocities, which are the highest tissue velocities routinely recorded, they are under 20 cm/s, and in diseased left ventricles the peaks of the S′ or e′ wave often are only a few cm/s in magnitude. In such signals, even small differences in how velocities are actually measured matter, because they lead to relatively high variations as a percentage of the absolute value. Tissue Doppler velocity–time displays, even though measuring reflectors in tissue, which compared with blood move with a relatively uniform velocity, produce a broad trace—they are rarely razor-thin. This so-called spectral broadening has several reasons: When using blood pool pulsed-wave Doppler to calculate volumetric flow rate, for example in the left ventricular outflow tract, it is recommended10 to trace the brightest signal within the signal envelope to calculate the time–velocity integral, which is then multiplied by the cross-sectional area of the flow to yield mL/beat. The rationale is that brightness, or signal strength, is proportional to the number of reflectors, and therefore, the brightest portion of the tracing indicates a weighted mean of the flow velocities present in the sample volume. This velocity is also called the ‘modal’ velocity (in descriptive statistics, the most frequently occurring value). Research comparing different measurement conventions for pulsed-wave Doppler tracings to calculate stroke volume has confirmed that the modal velocity is best suited to accurately calculate volumetric flow in the left ventricular outflow tract.11 The modal velocity is also less gain dependent than the velocities measured at the border of the visible envelope. Myocardium contains fibres with different orientations, which during the cardiac cycle change their internal orientation and the angle with the ultrasound beam. There is an apico-basal velocity gradient within the tissue. Several transducer elements are activated when the Doppler beam is generated, leading to a range of angles of insonation. Instrument-related factors such as gain, sample volume, and others. In tissue Doppler, however, the goal typically is not the calculation of a velocity–time integral, but the measurement of peak velocities, for example e′ or S′, as parameters of ventricular myocardial ‘function’. While measurement of the modal velocity has traditionally also been recommended for tissue Doppler,10 even guidelines sometimes show measurements taken at the outer border of the spectral envelope.12,13 However, it is not self-evident that the modal velocity, or the ‘middle’ of the tracing, necessarily is the best parameter to measure. One could instead argue that measuring the ‘leading edge’ of the Doppler tracing might be the easiest or most reproducible way to do this. In a historical analogy, one might remember that, in M-mode echocardiography, the leading-edge to leading-edge measurement convention for measuring aortic diameter has been adopted for reasons of reproducibility,14 in spite of erroneously incorporating the anterior aortic wall thickness into the measurement of the aortic lumen. In their study in this issue, Dhutia et al.15 analysed the accuracy of outer edge, middle line, and inner edge of tissue Doppler tracings both in vitro (comparing with an independent gold standard) and in patients (comparing with velocities calculated from M-mode and speckle tracking). They found that in vitro velocity values from the middle of the tracing corresponded best with the actual relative velocity of transducer vs. (stationary) tissue, and in patients they corresponded best with velocity calculated either by M-mode (dividing the excursion of the mitral annulus by the corresponding time interval) or by speckle tracking. Thus, measurements of the outer edge of the tracing, which are the most commonly used, are less accurate (although perhaps better reproducible) than those taken from the mid-thickness of the tracing. Some weaknesses in the methodology should be noted. The in vitro set-up of the study, where the tissue phantom is stationary, and the transducer is moved back and forth, reverses the situation in vivo and thus does not accurately replicate the ultrasound characteristics of tissue deformation. Furthermore, the true velocity in the in vitro set-up was obtained by ‘optical tracking’ (explained in their online appendix), the accuracy of which is unclear. Another important issue which this study brings up is the role of different manufacturers. Mårtensson et al.16 recently reported an in vitro comparison of the accuracy of six current commercially available ultrasound systems for tissue Doppler-based velocity, displacement, strain, and strain rate measurements. For velocity, they found mean differences between measured and true velocities which ranged between 0 ± 0.4 cm/s (corresponding to 0 ± 5% of the maximal value) in the most accurate ultrasound machine and 2.9 ± 0.4 cm/s (34 ± 3% of the maximal value) in the least accurate. Such systematic errors cannot be considered trivial and are a source of concern. Differences between measurements and true values for strain and strain rate were even higher, and in addition, there were timing errors. In the present study, four different machines from four manufacturers were applied to the in vitro setting. Surprisingly, Dhutia et al. report that ‘Using the in vitro model, the M-mode, speckle tracking, and tissue Doppler (middle line) traces broadly agreed with non-ultrasound optical assessment’ (the latter being the gold standard for motion in their in vitro set-up). How to reconcile the two studies? A closer look at the data in Table 1 of15 shows that the agreement in the current study was actually not substantially better than in the above-cited study.16 Biases in the modal velocity data generated by the different machines ranged from −1.72 to +0.4 cm/s, well in line with the mean errors reported by Mårtensson et al.16 Thus, important lessons from the study of Dhutia et al. are that More research is needed into the reproducibility of modal velocity measurements, and the findings from this study need to be confirmed with a more realistic tissue phantom capable of deformation. Furthermore, efforts need to be made to identify and remedy machine-related biases in such measurements. Therefore, it would be premature to call for new normal values and cut-offs for tissue Doppler based on the present study. Nevertheless, the study of Dhutia et al. commendably draws attention to a lack of consistency in current practice and should, at the very least, lead to being more clear in clinical practice as well as in research how tissue Doppler velocities actually are measured. the modal velocity, which can be approximated as the mid-line of the Doppler envelope thickness, is closest to the true mean tissue velocity in the sample volume; considerable systematic errors exist in tissue Doppler measurements.
No takes yet. Share an insight, caveat, or question.
Flachskampf et al. (2014) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: