SignificanceNon-invasive optical measurements of deep tissue (e.g., muscle) need to take into account confounding contributions from baseline and dynamic optical properties of superficial tissue (adipose tissue).AimDiscriminate superficial and deep tissue hemodynamics using data collected with frequency-domain (FD) near-infrared spectroscopy (NIRS) in a dual-slope (DS) configuration.ApproachExperimental data were collected in vivo on the forearm of three human subjects during a 3-min arterial occlusion or 1-min venous occlusion. Theoretical data were generated using diffusion theory for two-layered media with varying values of the reduced scattering coefficient (μs′) (range: 0.5 to 1.1 mm − 1) and absorption coefficient (μa) (range: 0.005 − 0.015 mm − 1) of the two layers, and top layer thickness (range: 2 to 8 mm). Data were analyzed using diffusion theory for a homogeneous semi-infinite medium.ResultsExperimental data in vivo were consistent with simulated data for a two-layered medium with a larger μs′ in the top layer, comparable absorption changes in the top and bottom layers during venous occlusion, and smaller absorption changes in the top vs. bottom layers during arterial occlusion.ConclusionsThe dataset generated by DS FD-NIRS may allow for discrimination of superficial and deep absorption changes in two-layered media, thus lending itself to individual measurements of hemodynamics in adipose and muscle tissue.