Key result
Free-breathing FURST yields similar myocardial ECV compared to conventional breath-held MOLLI.
Why the study?
MOLLI T1 mapping requires breath-holding and ECG gating with consistent R-R intervals, which is problematic for patients with AF, driving the need for a free-breathing and ungated framework.
Does free-breathing ungated FURST imaging provide comparable cardiac T1 and ECV maps to conventional MOLLI in human and canine subjects?
Cohort (n=24)
Does free-breathing ungated FURST imaging provide comparable cardiac T1 and ECV maps to conventional MOLLI in human and canine subjects?
Answers are grounded in this paper and cite their sources.
FURST provides comparable cardiac T1 and ECV mapping quality to MOLLI, offering a viable alternative for patients with atrial fibrillation or those unable to hold their breath.
Background Modified Look‐Locker imaging (MOLLI) T1 mapping sequences are acquired during breath‐holding and require ECG gating with consistent R‐R intervals, which is problematic for patients with atrial fibrillation (AF). Consequently, there is a need for a free‐breathing and ungated framework for cardiac T1 mapping. Purpose To develop and evaluate a free‐breathing ungated radial simultaneous multi‐slice (SMS) cardiac T1 mapping (FURST) framework. Study Type Retrospective, nonconsecutive cohort study. Population Twenty‐four datasets from 17 canine and 7 human subjects (4 males, years; 3 females, years). Canines were from studies involving AF induction and ablation treatment. The human population included separate subjects with suspected microvascular disease, acute coronary syndrome with persistent AF, and transthyretin amyloidosis with persistent AF. The remaining human subjects were healthy volunteers. Field Strength/Sequence Pre‐ and post‐contrast T1 mapping with the free‐breathing and ungated SMS inversion recovery sequence with gradient echo readout and with conventional MOLLI sequences at 1.5 T and 3.0 T. Assessment MOLLI and FURST were acquired in all subjects, and American Heart Association (AHA) segmentation was used for segment‐wise analysis. Pre‐contrast T1, post‐contrast T1, and ECV were analyzed using correlation and Bland–Altman plots in 13 canines and 7 human subjects. T1 difference box plots for repeated acquisitions in four canine subjects were used to assess reproducibility. The PIQUE image quality metric was used to evaluate the perceptual quality of T1 maps. Statistical Tests Paired t‐tests were used for all comparisons between FURST and MOLLI, with indicating statistical significance. Results There were no significant differences between FURST and MOLLI pre‐contrast T1 reproducibility ( and , ), FURST and MOLLI ECV ( and , ), or FURST and MOLLI PIQUE scores ( and , ). The ECV mean difference was with . Conclusions FURST had similar quality pre‐contrast T1, post‐contrast T1, and ECV maps and similar reproducibility compared to MOLLI.
No takes yet. Share an insight, caveat, or question.
Le et al. (2024) conducted a cohort in Healthy volunteers and cardiovascular disease (arrhythmia, suspected microvascular disease, transthyretin amyloidosis) (n=24). Free-breathing ungated radial simultaneous multi-slice cardiac T1 mapping (FURST) vs. Modified Look-Locker imaging (MOLLI) was evaluated on Myocardial extracellular volume (ECV) (MD 0.48, 95% CI 0.0006-0.96, p=0.05). FURST demonstrated similar myocardial extracellular volume (ECV) compared to conventional MOLLI.
Synapse has enriched 10 related papers. Consider them for comparative context: