Does the Edwards-Duromedics bileaflet aortic valve prosthesis differ from the St. Jude prosthesis in hydrodynamic characteristics in an in vitro model?
Does the Edwards-Duromedics bileaflet aortic valve prosthesis differ from the St. Jude prosthesis in hydrodynamic characteristics in an in vitro model?
In an in vitro model, the Edwards-Duromedics bileaflet valve demonstrated less regurgitation but a higher transvalvular pressure gradient and smaller orifice area compared to the St. Jude valve.
Answers are grounded in this paper and cite their sources.
In vitro hydrodynamic differences between bileaflet valves require clinical correlation; leaves open optimal prosthesis selection.
An in vitro comparison of the hydrodynamic characteristics of the Edwards-Duromedics (DM) and St Jude (SJ) bi-leaflet aortic valve prostheses is presented. Aortic valves 27 mm in diameter were mounted in a pulse duplicator simulating physiological pulsatile flow using a glycerol solution as the blood analogue fluid. Mean trans-valvular pressure difference (TPD) in systole, the per cent regurgitation (PCR) and the projected valve orifice area (VOA) were compared at a range of flow rates and heart rates to reflect the functioning of the valve under resting as well as exercise conditions. Our results showed that the TDP for DM (0.99 +/- 0.28 kPa) valve was slightly larger than the SJ (0.81 +/- 0.19 kPa) valve (mean difference of 0.18 +/- 0.26 kPa, P less than 0.05). However, the PCR for the DM (13.6 +/- 7.8) valve was smaller than the SJ (17.9 +/- 9.0) valve (mean difference of 4.3 +/- 4.2%, P less than 0.01). Moreover, VOA for the DM (2.51 +/- 0.10 cm2) valve was smaller than the SJ (3.13 +/- 0.04 cm2) valve (mean difference of 0.63 +/- 0.10 cm3, P less than 0.001).
No takes yet. Share an insight, caveat, or question.
Chandran et al. (1988) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: