Key result
St. Jude bileaflet aortic valve links to higher performance index than porcine valves across sizes.
Why the study?
The St. Jude bileaflet disc cardiac valve prosthesis was newly designed to avoid the hemodynamic drawbacks of other prostheses, motivating an evaluation of its in vivo flow characteristics compared with other commonly used aortic prostheses.
Does the St. Jude bileaflet aortic valve improve in vivo hydrodynamic characteristics compared to Björk-Shiley, Hancock, and Carpentier-Edwards prostheses in patients undergoing aortic valve replacement?
Observational (n=96)
Does the St. Jude bileaflet aortic valve improve in vivo hydrodynamic characteristics compared to Björk-Shiley, Hancock, and Carpentier-Edwards prostheses in patients undergoing aortic valve replacement?
The St. Jude bileaflet aortic valve demonstrates superior early postoperative in vivo hemodynamic performance (higher performance index, larger effective orifice area, and lower systolic gradient during flow augmentation) compared to Hancock and Carpentier-Edwards porcine bioprostheses.
Answers are grounded in this paper and cite their sources.
Offers hydrodynamic insights for aortic prosthesis selection; leaves open correlation with clinical outcomes.
The St. Jude valve is a new bileaflet disc cardiac valve prosthesis designed to avoid some of the hemodynamic drawbacks of other prostheses. The in vivo flow characteristics of the St. Jude aortic valve (42 patients) were studied and compared with those of three other commonly used aortic prostheses. Björk-Shiley (12 patients), Hancock (27 patients) and Carpentier-Edwards (15 patients). The studies, performed 24 to 48 hours after surgery, included measurements at rest and during augmentation of valve flow by infusion of isoproterenol. The mean performance index for valves of all sizes is higher for the St. Jude than for either porcine valve, both at rest and during isoproterenol infusion (p less than 0.05). Utilizing data both at rest and with isoproterenol, the relation of valve flow and mean systolic gradient for each size of St. Jude valve (19 to 25 mm) indicates the occurrence of small increases in gradient (5.3 to 8.2 mm Hg) as valve flow increases, ranging from 161 to 436 ml/systolic X min. A direct comparison of valve flow and gradient data for all size 25 and 23 mm prostheses at rest indicates a tendency for a lower mean systolic gradient in both mechanical valves than in either porcine valve (p = 0.07). With isoproterenol augmentation of valve flow in 25 mm valves, the gradient is less (p less than 0.05), and the effective orifice area and performance index are larger (p less than 0.05) for the St. Jude than for either porcine valve.(ABSTRACT TRUNCATED AT 250 WORDS)
No takes yet. Share an insight, caveat, or question.
Gray et al. (1984) conducted an observational in Aortic valve replacement (n=96). St. Jude bileaflet aortic valve prosthesis vs. Björk-Shiley, Hancock, and Carpentier-Edwards aortic prostheses was evaluated on In vivo hemodynamic characteristics (mean performance index, mean systolic gradient, and effective orifice area) (p=<0.05). The St. Jude bileaflet aortic valve showed a higher mean performance index across all sizes than Hancock or Carpentier-Edwards porcine valves at rest and during isoproterenol infusion.
Synapse has enriched 10 related papers. Consider them for comparative context: