Why the study?
Does defibrillator lead system revision restore a 10-J safety margin in patients with a rise in chronic defibrillation energy requirements?
Does defibrillator lead system revision restore a 10-J safety margin in patients with a rise in chronic defibrillation energy requirements?
In patients with a progressive increase in chronic defibrillation energy requirements, lead system revision successfully restores an adequate safety margin, highlighting the need for routine DER reevaluation.
Answers are grounded in this paper and cite their sources.
Lead revision may restore margins in rising DER; leaves open whether routine reevaluation improves outcomes.
The chronic defibrillation energy requirement (DER) is believed to remain clinically stable in patients with defibrillators. Six patients (two with an epicardial and four with a nonthoracotomy system) were identified with a rise in their chronic DER, which eliminated a 10-J safety margin, thus necessitating a defibrillator lead system revision. The mean increase in DER was 14.7 +/- 4 J and was discovered at a mean of 16.0 +/- 18 months (range 2-41) following implantation. Management included placement of a defibrillator with a biphasic waveform, placement of an additional defibrillation electrode, or both. At 2 months following revision of the defibrillation system, a 10-J DER safety margin was present in each patient. In some patients, there is a progressive increase in the chronic DER with elimination of a 10-J safety margin necessitating revision of the defibrillation system. Routine reevaluation of the chronic DER, therefore, is necessary to identify these patients.
No takes yet. Share an insight, caveat, or question.
Daoud et al. (1997) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: