Describes a mechanism for a pseudo-endless loop tachycardia in a DDD pacemaker caused by VPC interference and upper rate limit behavior rather than retrograde conduction.
Answers are grounded in this paper and cite their sources.
Clinicians should consider VPC interference with upper rate behavior in apparent endless loop tachycardia; case reports leave open true incidence.
A 3-beat episode of ventricular arrhythmia was recorded in a patient with an AV universal (DDD) pacemaker. This arrhythmia mimicked "endless loop" tachycardia (ELT) because it appeared following a ventricular premature contraction (VPC), its rate approached the preset upper-rate limit and the regular P waves disappeared during the episode. However, as retrograde conduction was absent ELT could not exist. The mechanism of this arrhythmia was based on a combination of interference by VPC of the regular sequence of QRS complexes produced by P wave and the pacemaker commitment to maintain an upper rate limit by prolongation of the programmed pacemaker AV delay.