Why the study?
Does the heparin sensitivity of different lots of APTT reagents from the same manufacturer vary significantly?
Does the heparin sensitivity of different lots of APTT reagents from the same manufacturer vary significantly?
Significant lot-to-lot variation in heparin sensitivity of APTT reagents from the same manufacturer can lead to inconsistent heparin dosing if standard ratios are used.
Answers are grounded in this paper and cite their sources.
May warrant lot-specific APTT calibration in labs to avoid dosing variability; leaves open standardization needs for heparin monitoring.
The variations between different lots of activated partial thromboplastin time (APTT) produced by three major North American suppliers were evaluated over the past eight years. The authors found significant variations between the heparin sensitivity of the APTT reagents produced under the same name by the same supplier. The variations were so much that, using the recommended APTT ratio or prolongation of APTT for monitoring heparin therapy, one would have achieved significantly different intensity of heparinization from year to year.
No takes yet. Share an insight, caveat, or question.
Shojania et al. (1988) studied this question.