Why the study?
Are home blood pressure monitors accurate compared to 2-observer auscultation when used by device owners?
Are home blood pressure monitors accurate compared to 2-observer auscultation when used by device owners?
Most home blood pressure devices used by their owners are not accurate to within 5 mm Hg compared to standard auscultation, highlighting the need to verify device-owner pairing accuracy.
Answers are grounded in this paper and cite their sources.
Home BP devices often inaccurate on owners; leaves open need for individualized validation before guiding therapy.
OBJECTIVE: To examine the accuracy of home blood pressure (BP) devices, on their owners, compared to auscultatory reference standard BP measurements. METHODS: Eighty-five consecutive consenting subjects ≥18 years of age, who owned an oscillometric home BP device (wrist or upper-arm device), with BP levels between 80-220/50-120 mm Hg, and with arm circumferences between 25-43 cm were studied. Pregnancy and atrial fibrillation were exclusion criteria. Device measurements from each subject's home BP device were compared to simultaneous 2-observer auscultation using a mercury sphygmomanometer. Between-group mean comparisons were conducted using paired t-tests. The proportion of patients with device-to-auscultatory differences of ≥5, 10, and 15 mm Hg were tabulated and predictors of systolic and diastolic BP differences were identified using linear regression. RESULTS: Mean age was 66.4 ± 11.0 years, mean arm circumference was 32.7 ± 3.7 cm, 54% were female and 78% had hypertension. Mean BPs were 125.7 ± 14.0/73.9 ± 10.4 mm Hg for home BP devices vs. 129.0 ± 14.7/72.9 ± 9.3 for auscultation (difference of -3.3 ± 7.3/0.9 ± 6.1; P values <0.0001 for systolic and 0.17 for diastolic). The proportion of devices with systolic or diastolic BP differences from auscultation of ≥5, 10, and 15 mm Hg was 69%, 29%, and 7%, respectively. Increasing arm circumference was a statistically significant predictor of higher systolic (parameter estimate 0.61 per cm increase; P value 0.004) and diastolic (0.38; 0.03) BP. CONCLUSIONS: Although mean differences from 2-observer auscultation were acceptable, when tested on their owners, most home BP devices were not accurate to within 5 mm Hg. Ensuring acceptable accuracy of the device-owner pairing should be prioritized.
No takes yet. Share an insight, caveat, or question.
Ringrose et al. (2017) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: