This is the authors' abstract. We don't add key points for this paper.
The paper details the conversion of an existing isothermal bomb calorimeter to an adiabatic calorimeter with automatic control. Thermistors in the inner and outer vessels are included in two arms of an a.c. Wheatstone bridge: any rise in temperature of the inner vessel above that of the outer vessel results in phase reversal of the output voltage from the bridge, which operates a gas-filled tetrode controlling a 500 W heater in the outer vessel. A second tetrode, operated by off-balance voltage, controls a 1000 W heater in the outer vessel which accommodates the initial rapid temperature rise when the bomb is fired. The equipment has been found to be more rapid and much less tedious for routine use than the usual isothermal method. Determinations of heat content take approximately thirty minutes: the coefficient of variation of a single determination, with an average temperature rise of 1.9° C, is 0.34%.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Raymond et al. (1957) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: