Why the study?
LBBAP is typically performed with a lumenless lead, but stylet-driven leads were recently introduced, prompting a comparison of acute and chronic outcomes between the two lead and delivery systems.
Does a stylet-driven lead system provide comparable procedural success and conduction system capture maintenance to a lumenless lead system for left bundle branch area pacing in patients with standard pacing indications?
Does a stylet-driven lead system provide comparable procedural success and conduction system capture maintenance to a lumenless lead system for left bundle branch area pacing in patients with standard pacing indications?
Stylet-driven leads offer a viable alternative to lumenless leads for LBBAP with comparable acute success, though longer follow-up QRS duration suggests potential loss of conduction system capture in some patients.
Answers are grounded in this paper and cite their sources.
SDL with CPS Locator shows comparable LBBAP success to LLL; leaves open equivalence in prospective randomized trials.
Introduction Left bundle branch area pacing (LBBAP) is typically performed using the lumenless (LLL) Medtronic SelectSecure TM Model 3830 pacing lead. Stylet‐driven leads (SDL) for LBBAP were recently introduced. Objective To compare the Abbott Tendril STS TM 2088TC SDL and CPS Locator TM 3D delivery system to the Medtronic LLL and delivery system acutely and chronically. Methods Fifty consecutive patients who received SDL using the CPS Locator TM 3D delivery system for standard pacing indications were included in this study. A comparison was made with 229 LLL patients from a retrospective database, evaluating baseline characteristics, procedural outcomes, lead and system performance, and follow‐up data. Results LBBAP was successful in 84% of SDL and 79% of LLL patients ( p = 0.55), with no significant differences in procedural characteristics or complications. Minor pacing differences were noted. CPS Locator TM 3D performed similarly to C315His® delivery system with a trend for less lead placement attempts. Conduction system capture (CSC) pattern at implant did not differ between both groups. CSC maintenance was similar between SDL and LLL. However, follow‐up QRS duration was significantly longer in SDL group (154 ms vs. 129 ms, p = 0.004), suggesting a possibility of CSC loss in some patients. Conclusion SDL is a viable alternative to LLL, demonstrating comparable acute and midterm pacing parameters. CPS Locator TM 3D delivery system showed comparable performance and an excellent safety profile. Acute CSC with SDL and LLL is similar, but its long‐term maintenance will require further study.