Key result
Left bundle branch pacing using lumenless leads resulted in shorter fluoroscopy time compared to stylet-driven leads, with comparable complication rates.
Why the study?
The authors sought to describe their experience with left bundle branch pacing using lumenless leads compared to stylet-driven leads.
Does left bundle branch pacing using lumenless leads compared to stylet-driven leads improve procedural or electrical outcomes in patients requiring pacing?
Cohort (n=124)
Yes
Does left bundle branch pacing using lumenless leads compared to stylet-driven leads improve procedural or electrical outcomes in patients requiring pacing?
Left bundle branch pacing using either lumenless or stylet-driven leads offers comparable success, safety, and paced QRS duration, though stylet-driven leads require longer fluoroscopy times.
Answers are grounded in this paper and cite their sources.
May shorten fluoroscopy time without added complications in left bundle branch pacing; leaves open optimal lead selection pending randomized data.
Conduction system pacing started with His bundle pacing (HBP) and then rapidly switched gears into left bundle branch pacing (LBBP). We describe our center’s experience with LBBP using either lumenless leads (LLLs) or stylet-driven leads (SDLs). Patients who were admitted to two tertiary centers between 1 April 2021 and 30 June 2024 and met the guidelines for pacing were recruited and prospectively followed up. A total of 124 patients underwent permanent pacemaker (PPM) implantation using the LBBP technique with a mean follow-up of 19.7 ± 13.3 months. In total, 90 patients were implanted with LLLs and 34 with SDLs. There was no significant difference in the procedural time and final paced QRS duration, but fluoroscopy time was significantly longer in the SDLs (26.2 ± 17.7 min vs. 17.5 ± 13.0 min, respectively, p = 0.026). The on-table impedance values were also significantly higher in the LLLs, and this persisted throughout the follow-up. There were no differences in the rates of complications. The success of conduction system pacing implantation with SDLs and LLLs is comparable with reasonable safety and reliable outcomes. Good pre-implant patient selection will contribute to improved outcomes.
No takes yet. Share an insight, caveat, or question.
Teo et al. (2025) conducted a cohort in Indication for permanent pacemaker (n=124). Left bundle branch pacing using lumenless leads (LLLs) vs. Left bundle branch pacing using stylet-driven leads (SDLs) was evaluated on Fluoroscopy time (minutes) (p=0.026). Left bundle branch pacing using lumenless leads resulted in shorter fluoroscopy time compared to stylet-driven leads, with comparable complication rates.