Why the study?
The use of CRT-D devices with LV sensing has created controversy regarding programming parameters, particularly left ventricular T wave protection designed to prevent pacing stimuli during the LV vulnerable period.
How should left ventricular T wave protection (LVTP) and desynchronization-detection algorithms be programmed in CRT-D devices with left ventricular sensing?
How should left ventricular T wave protection (LVTP) and desynchronization-detection algorithms be programmed in CRT-D devices with left ventricular sensing?
Appropriate programming of LVTP in Biotronik CRT-D devices requires weighing the prevention of episodic desynchronization against the preservation of valuable diagnostic data.
Answers are grounded in this paper and cite their sources.
LVTP deactivation may prevent desynchronization in Biotronik CRT-D; leaves open optimal programming pending prospective trials.
The use of CRT-D devices with left ventricular (LV) sensing has created controversy about programming various parameters especially the left ventricular T wave protection (LVTP) designed to prevent the delivery of a pacing stimulus into the LV vulnerable period. Such devices are available from two manufacturers. This review focuses only on those provided by Biotronik. As the LVTP controls LV sensing, some investigators have advocated turning off the LVTP to prevent episodic desynchronization known a CRT pacing interrupt. However, LVTP off reduces but does not eliminate this type of desynchronization if triggering of an LV stimulus upon right ventricular sensing (RVs) is programmed on. Deactivation of the LVTP incurs loss of diagnostic data provided by CRT pacing interrupt itself. By choice, the occurrence of CRT pacing interrupt can be totally eliminated by appropriate programming of the LV upper rate interval, LVTP and triggering of an LV pacing event upon RVs. Various programmability options are available according to clinical circumstances. As a rule, clinical judgement must weigh the potential diagnostic benefit of preserving the LVTP capable of recording of episodic CRT pacing interrupt against the loss of diagnostic benefit when LVTP is programmed off (with or without triggering of an LV stimulus upon RVs).
No takes yet. Share an insight, caveat, or question.
Barold et al. (2024) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: