Why the study?
Sense-B-noise is a newly reported potential cause of inappropriate S-ICD shocks with an unknown nature, and whether reprogramming to the secondary sensing vector can consistently solve this issue is unclear.
What is the incidence of sense-B-noise in S-ICD patients and is the secondary sensing vector available for reprogramming?
What is the incidence of sense-B-noise in S-ICD patients and is the secondary sensing vector available for reprogramming?
Sense-B-noise is a rare but relevant issue in S-ICD patients (3% incidence), and most vulnerable patients can be managed by reprogramming to the secondary sensing vector.
Answers are grounded in this paper and cite their sources.
Sense-B-noise may trigger S-ICD shocks; leaves optimal vector reprogramming open for prospective validation.
BACKGROUND: Sense-B-noise is a newly reported possible cause of inappropriate shocks in patients with subcutaneous implantable cardioverter-defibrillators (S-ICDs). The nature of that noise is unknown, it is not related to mechanical failure of the S-ICD system. Reprogramming to the secondary sensing vector is suggested by the producer as a possible solution. METHODS: We analyzed the medical records of S-ICD recipients from two university clinical centers (Gdansk and Szczecin, Poland). Our aim was to determine the rate of sense-B-noise, and whether the secondary sensing vector would be available for reprogramming if such a problem occurred in our patients. RESULTS: The sense-B-noise issue affected three patients in our cohort (3%), which corresponds to the incidence of 0.012 events per patient-year of follow-up. The primary vector was permanently used in 47 patients (52%), secondary in 28 (31%), and alternate in 16 (17%), respectively. Therefore, the total number of patients potentially vulnerable to sense-B noise (with the primary or alternate vector programmed permanently) was 63 (69%). Among those 63 patients, 51 individuals (81%) had also the secondary vector available for permanent use. CONCLUSION: The sense-B-noise affected 3% of patients in our cohort, with an incidence of 0.012 per patient-year of follow-up. Most patients potentially vulnerable to sense-B noise could be reprogrammed to the secondary sensing vector, if necessary. Further investigation of the sense-B-noise issue is needed.
No takes yet. Share an insight, caveat, or question.
Budrejko et al. (2023) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: