The developed two-step immunoradiometric assay provides a rapid and precise method for measuring N-terminal proANP, a marker of heart failure, in human plasma without prior extraction.
Answers are grounded in this paper and cite their sources.
Facilitates precise NT-proANP measurement in plasma; leaves open its diagnostic role in heart failure pending clinical validation.
Recently, the N-terminal fragment of proatrial natriuretic peptide (N-terminal proANP) has been proposed as a marker of chronic congestive heart failure. In this study, we established a two-step immunoradiometric assay using monoclonal antibodies and synthetic N-terminal proANP (1–67) as a standard. It allows us to measure plasma N-terminal proANP in only 4 h without prior extraction. The detection limit of this assay was 15 pmol/L for a 100 μL sample of plasma. Within-run CVs ranged from 1.7% to 2.9% and between-run CVs ranged from 4.2% to 5.1%. The dilution curves of plasma samples showed good linearity and analytical recovery was 89–104%. The mean (±SD) N-terminal proANP in plasma of 33 healthy subjects was 188 (±71) pmol/L and 1030 (±411) pmol/L in 25 patients with heart failure. Our immunoradiometric assay is rapid and precise enough for routine determination of N-terminal proANP in human plasma.
No takes yet. Share an insight, caveat, or question.
Numata et al. (1998) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: