Why the study?
Does coronary sinus shocking lead placement achieve acceptable defibrillation thresholds in patients with advanced cardiomyopathy and high DFTs?
Does coronary sinus shocking lead placement achieve acceptable defibrillation thresholds in patients with advanced cardiomyopathy and high DFTs?
Coronary sinus shocking lead placement may serve as a salvage option for patients with advanced heart failure and unacceptably high defibrillation thresholds during ICD implantation.
Answers are grounded in this paper and cite their sources.
May serve as salvage option in refractory high-DFT ICD cases; leaves open need for prospective validation.
BACKGROUND: We report a series of three patients whose implantable cardioverter-defibrillators (ICD) implants were unsuccessful due to inability to achieve defibrillation thresholds (DFT) at maximum available energy after failure of standard modification and enhancement procedures. All patients had advanced cardiomyopathy. METHODS: Use of the coronary sinus (CS) for left ventricular (LV) shocking electrode placement resulted in acceptable DFTs in each patient. The position of the shocking coil in all three patients was posterior, and in two patients alongside a left ventricular CS pacing lead. The best shocking configuration tested was LV (CS) + CAN (Anode) to RV (cathode) in each patient. The short- and long-term outcomes of these patients is presented and discussed. CONCLUSION: This approach is suggested as a salvage option for those problematic patients who have unacceptable DFT results at implantation of an endovascular ICD system.
No takes yet. Share an insight, caveat, or question.
Faheem et al. (2010) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: