The newly designed AC amplifier demonstrates high common mode rejection ratio and dynamic range, potentially improving the quality of electrophysiological recordings in noisy environments.
May improve signal quality in noisy cardiac recordings; leaves open clinical validation before adoption.
We present the design, simulation and test results of a new AC amplifier for electrophysiological measurements based on a three op-amp instrumentation amplifier (IA). The design target was to increase the common mode rejection ratio (CMRR), thereby improving the quality of the recorded physiological signals in a noisy environment. The new amplifier actively suppresses the DC component of the differential signal and actively reduces the common mode signal in the first stage of the IA. These functions increase the dynamic range of the amplifier's first stage of the differential signal. The next step was the realization of the amplifier in a single chip technology. The design and tests of the new AC amplifier with a differential gain of 79.2 dB, a CMRR of 130 dB at 50 Hz, a high-pass cutoff frequency at 0.01 Hz and common mode reduction in the first stage of the 49.8 dB are presented in this paper.