This is the authors' abstract. We don't add key points for this paper.
We analyze the input impedance and CMRR requirements for an amplifier for bioimpedance measurements when considering the capacitive components of the electrode-skin contact impedance. We describe an ac-coupled instrumentation amplifier (IA) that, in addition to fulfilling those requirements, both provides interference and noise reduction, and yields a zero phase shift over a wide frequency band without using broadband op amps.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Pallás-Areny et al. (1993) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: